Upgrading the Program
In this section, the procedure to upgrade the program is explained.
The DS-7001 system bedside monitor adopts the compact flash card (hereinafter referred to as
CF card) for the upgrading procedure.
There are following 2 methods to upgrade the program.
Standard Install Process
This is the standard install process. The upgrading can be performed from the preset menu.
For normal usage, use this method to upgrade the DS-7001 system bedside monitor.
CAUTION
During the standard install process, do not perform the following
actions. Otherwise, the software error will occur, and forced install
process may become necessary.
Inserting / removing of the CF card
Plugging / unplugging of the AC power cable
Forced Install Process
This process forcibly installs or upgrades the program to the DS-7001 system bedside monitor.
The install process will automatically start when the system is started. Use this forced install
process when failure of the DS-7001 system bedside monitor occurs, or when the standard
install process from the preset menu is not possible.
CAUTION
For normal usage, the forced install process will not be necessary.
Use this process only when the standard install process is not
possible due to software error.
4
-
7
Содержание DS-7001
Страница 28: ...xxvi Blank Page ...
Страница 40: ...1 6 Blank Page ...
Страница 46: ...2 6 Blank Page ...
Страница 57: ... SV 900 SV 300 Servo i Servo s CJ 501 Servo i Servo s CJ 502 SV 900 SV 300 CJ 500 3 11 ...
Страница 78: ...4 16 Blank Page ...
Страница 104: ...5 26 Blank Page ...
Страница 114: ...6 10 Reference For details of the battery lamp refer to 3 Preparation Before Use To Turn ON the Power ...
Страница 126: ...7 12 Blank Page ...
Страница 150: ...9 14 Blank Page ...
Страница 151: ...Printed in Japan 4R2160 200606 39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 ...