Analog Board (PCB-6972)
The analog board directly connects to the PCB-6940 CPU board.
It is composed of ECG / respiration measurement circuit, and converts analog signal to digital
signal, and outputs to the CPU board. The measurement is controlled by the CPU board.
Also, it is connected to the NIBP amplifier board, SpO
2
unit, and the measurement data is output
to the CPU board via serial communication as with the signals from each measurement circuit.
Block Diagram of the Analog Board
A/D
Converter
ECG/RESP
Measurement Circuit
Controller
MPX
SpO2
Measurement Circuit
NIBP Sensor Board
Isolation Area
ECG
Connector
Cuff
Connector
SpO2
Connector
CPU
Board
CPU
Board
CPU
Board
ECG/
RESP
NIBP
SpO2/PR
Control Signal
Control Signal
Control
Signal
9
-