Equipment Status Alarm Message
The equipment status alarm message will be displayed when proper monitoring cannot be
performed. Some messages are displayed on the numeric data display area.
The message will be displayed according to the priority.
If more than one alarm are generated, the two higher priority alarms will be alternately
displayed.
CAUTION
While the “LEAD OFF” message is displayed, the HR alarm will not
generate. Leaving this condition unresolved may result in missing a
sudden change of the patient. Promptly check the electrodes when
this message is displayed.
Priority Message
Displayed
Color
Description
DEMO
DEMO
mode
High BATTERY
The remaining battery capacity is less than 10 minutes.
TELE. ER
Telemetry module error.
NO CARD
The CF card is not inserted.
Com Err
Communication error with the SpO
2
hardware
CVA
ECG waveform is superimposed on the respiration
waveform.
Sensor
Err
SpO
2
sensor failure.
PULSE?
Searching
pulse.
MOTION
Body motion is present during SpO
2
measurement.
NIBP ERR
NIBP measurement error.
(Cuff inflation pressure, measurement duration has
exceeded the upper limit.)
CF ERR
Detected error during CF card usage.
CF
FULL
Yellow
No space on the CF card to write the data.
CF REC
Recording waveform data on the CF card.
Low CF
INIT
Blue
Checking the inserted CF card.
7
-
10
Содержание DS-7001
Страница 28: ...xxvi Blank Page ...
Страница 40: ...1 6 Blank Page ...
Страница 46: ...2 6 Blank Page ...
Страница 57: ... SV 900 SV 300 Servo i Servo s CJ 501 Servo i Servo s CJ 502 SV 900 SV 300 CJ 500 3 11 ...
Страница 78: ...4 16 Blank Page ...
Страница 104: ...5 26 Blank Page ...
Страница 114: ...6 10 Reference For details of the battery lamp refer to 3 Preparation Before Use To Turn ON the Power ...
Страница 126: ...7 12 Blank Page ...
Страница 150: ...9 14 Blank Page ...
Страница 151: ...Printed in Japan 4R2160 200606 39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 ...