Battery Pack
The battery pack needs to be replaced when used for 2 years or charged / discharged for 300
times. To ensure reliability of safety, function, and performance of the DS-7001 system
bedside monitor, the battery pack must be replaced periodically. Otherwise, it may cause the
following degradation.
Decrease in battery operation time or battery failure
CAUTION
The periodic replacement parts must be replaced within specified
period.
1 Turn OFF the power, and detach the bottom panel.
Remove the 2 screws (3
´
8W sems) shown in circle,
and detach the bottom panel.
2 Remove the battery.
Disconnect the battery from the cable,
and reconnect a new battery.
Place the battery inside the
compartment, and attach the bottom
panel. After the assembly, connect
the AC power cable, and check that
the battery lamp lights. The battery
lamp will light in green when fully
charged, and light in orange during
charging.