Precautions for Installing the Equipment
The following environmental conditions should be observed when operating the DS-7001.
・
Surrounding Temperature : 10
~
40
°
C
・
Relative Humidity
: 30
~
85% (non-condensing)
The DS-7001 is intended for patient monitoring in the general ward, ICU, CCU, OR and during
in-hospital transportation. Do not use in the MRI environment, inside hyperbaric chamber,
flammable anesthetic gas and outside hospital (including inside ambulance).
The power source should fulfill the following condition.
・
Always use a hospital grade, 3 prong grounded power cable.
・
Verify the power voltage and frequency before connecting to an AC power source.
・
Use the power source that can provide adequate power to the device.
Pay attention when installing or storing the device. Do not install or store in the following
locations.
・
Where chemicals are stored or gas may generate
・
Where the equipment will be subject to splashing water or humidity from a nebulizer or
vapolizer
・
Where the equipment will be subject to direct sunlight
・
Unstable place with inclination, vibration, or shock
Proper ventilation is required to cool the device.
・
Do not cover the vent hole located at the rear side of the monitor.
A minimum space of 3 cm is required between the monitor and the wall.
・
If the monitor is embedded in a wall or surrounded by a wall, a minimum space of 3 cm is
required above the monitor.
W ARNI NG
l
Do not cover the vent hole located at the rear side of the monitor.
l
If this device is used under adverse environmental conditions, not
only that the device cannot deliver its maximum performance, the
device may be damaged and safety cannot be ensured.
3
-
2
Содержание DS-7001
Страница 28: ...xxvi Blank Page ...
Страница 40: ...1 6 Blank Page ...
Страница 46: ...2 6 Blank Page ...
Страница 57: ... SV 900 SV 300 Servo i Servo s CJ 501 Servo i Servo s CJ 502 SV 900 SV 300 CJ 500 3 11 ...
Страница 78: ...4 16 Blank Page ...
Страница 104: ...5 26 Blank Page ...
Страница 114: ...6 10 Reference For details of the battery lamp refer to 3 Preparation Before Use To Turn ON the Power ...
Страница 126: ...7 12 Blank Page ...
Страница 150: ...9 14 Blank Page ...
Страница 151: ...Printed in Japan 4R2160 200606 39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 ...