Compliance to the Electromagnetic Immunity (2)
The DS-7001 is intended for use in the electromagnetic environment specified below. It should
be assured that the DS-7001 is used in such an environment.
Immunity
Test
IEC60601-1-2
Test Level
Complianc
e Level
Electromagnetic Environment - Guidance
Conducted RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
3Vrms
150kHz
~
80MHz
3V/m
80MHz
~
2.5GHz
3Vrms
3V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the DS-7001, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.
Recommended Separation Distance
d = 1.2
P
d = 1.2
P
80MHz
~
800MHz
d = 2.3
P
800MHz
~
2.5GHz
Where P is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in
each frequency range.
Interference may occur in the vicinity of
equipment marked with the following symbol:
Note 1 : At 80Mhz and 800MHz, the higher frequency range applies.
Note 2 : These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast can not be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic
site survey should be considered. If the measured field strength in the location in which this
monitor should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating this monitor.
b
Over the frequency range 150kHz to 80MHz, field strength should be less than 3V/m.
xxiv
Содержание DS-7001
Страница 28: ...xxvi Blank Page ...
Страница 40: ...1 6 Blank Page ...
Страница 46: ...2 6 Blank Page ...
Страница 57: ... SV 900 SV 300 Servo i Servo s CJ 501 Servo i Servo s CJ 502 SV 900 SV 300 CJ 500 3 11 ...
Страница 78: ...4 16 Blank Page ...
Страница 104: ...5 26 Blank Page ...
Страница 114: ...6 10 Reference For details of the battery lamp refer to 3 Preparation Before Use To Turn ON the Power ...
Страница 126: ...7 12 Blank Page ...
Страница 150: ...9 14 Blank Page ...
Страница 151: ...Printed in Japan 4R2160 200606 39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 ...