Precautions about the Maintenance
Safety Inspection and Maintenance
For safe operation of the equipment, regular inspection and maintenance is required. Once a
year, check all cables, devices, and accessories for damage, earth impedance, earth and
leakage currents, and all alarm functions. Also, ensure that all safety labels are legible.
Maintain a record of these safety inspections.
Immediate maintenance has to be carried out if ;
·
the equipment was subjected to extreme mechanical stress, e.g. after a heavy fall.
·
the equipment was subjected to liquid spill.
·
the monitoring function is interrupted or disturbed.
→
“4. Maintenance”
·
parts of the equipment enclosure are cracked, removed, or lost.
·
any connector or cable shows signs of deterioration.
WARNING
Never open the housing while the equipment is in operation or
connected to hospital grade outlet as it may result in electric shock.
Maintenance, Modifications, and Repairs
Fukuda Denshi is liable for the safety, reliability, and performance of its equipment only if;
·
Maintenance, modifications, and repairs are carried out by authorized personnel.
·
Components are used in accordance with Fukuda Denshi operating instructions.
A full technical description of the DS-7001 is available from your local Fukuda Denshi
representative.
Precautions about the Pacemaker
WARNING
Minute ventilation rate-adaptive implantable pacemakers can
occasionally interact with certain cardiac monitoring and diagnostic
equipment, causing the pacemakers to pace at their maximum
programmed rate. The cardiac monitoring and diagnostic equipment
may possibly send wrong information.
If such event occurs, please disconnect the cardiac monitoring and
diagnostic equipment, or follow the procedures described in the
operation manual of the pacemaker.
(For more details, contact FUKUDA DENSHI personnel, your
institution’s professionals, or your pacemaker distributors.)
Reference
“Minute Ventilation Rate-Adaptive Pacemakers”
FDA alerts health professionals that minute ventilation
rate-adaptive implantable pacemakers can occasionally
interact with certain cardiac monitoring and diagnostic
equipment, causing pacemakers to pace at their maximum
programmed rate.
[October 14, 1998 (Letter: www.fda.gov/cdrh/safety.html) -
FDA]
viii
Содержание DS-7001
Страница 28: ...xxvi Blank Page ...
Страница 40: ...1 6 Blank Page ...
Страница 46: ...2 6 Blank Page ...
Страница 57: ... SV 900 SV 300 Servo i Servo s CJ 501 Servo i Servo s CJ 502 SV 900 SV 300 CJ 500 3 11 ...
Страница 78: ...4 16 Blank Page ...
Страница 104: ...5 26 Blank Page ...
Страница 114: ...6 10 Reference For details of the battery lamp refer to 3 Preparation Before Use To Turn ON the Power ...
Страница 126: ...7 12 Blank Page ...
Страница 150: ...9 14 Blank Page ...
Страница 151: ...Printed in Japan 4R2160 200606 39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 ...