EMC Guidance
This equipment complies with IEC60601-1-2 (2001). However, if portable transmitter or
wireless LAN equipment is used in close proximity, the electromagnetic influence may largely
exceed the compliance level and may cause unexpected phenomenon such as noise
interference on the waveform, etc.
Therefore, this equipment should be used in a location specified by each medical institution.
If any unexpected noise interference on the waveform or failure to the peripheral device occurs,
stop using the equipment and follow the instruction of the technician.
The following is the information relating to EMC (Electromagnetic Compatibility).
(When using this equipment, verify that it is used within the environment specified below.)
Compliance to the Electromagnetic Emissions
The DS-7001 is intended for use in the electromagnetic environment specified below.
Emissions Test Compliance
Electromagnetic Environment - Guidance
RF Emissions
CISPR 11
Group 1
The equipment uses RF energy that is necessary for the
internal functioning of the equipment itself. Therefore, its
RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class A
This equipment is suitable for use in all establishments
other than domestic and those directly connected to a
low-voltage power supply network which supplies buildings
used for domestic purpose.
xxii
Содержание DS-7001
Страница 28: ...xxvi Blank Page ...
Страница 40: ...1 6 Blank Page ...
Страница 46: ...2 6 Blank Page ...
Страница 57: ... SV 900 SV 300 Servo i Servo s CJ 501 Servo i Servo s CJ 502 SV 900 SV 300 CJ 500 3 11 ...
Страница 78: ...4 16 Blank Page ...
Страница 104: ...5 26 Blank Page ...
Страница 114: ...6 10 Reference For details of the battery lamp refer to 3 Preparation Before Use To Turn ON the Power ...
Страница 126: ...7 12 Blank Page ...
Страница 150: ...9 14 Blank Page ...
Страница 151: ...Printed in Japan 4R2160 200606 39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 ...