CAUTION
Patient Admit / Discharge
·
If you start monitoring a new patient without performing a discharge
procedure for the previous patient, new data will be added to the previous
data which will result in inaccuracy. Before monitoring, make sure the
current patient information is suitable for the patient’s condition.
CF Card
·
Use only the specified IC card (FCF-32/FCF-64/FCF-128).
·
When recording the waveform data on the CF card, use the CF card
formatted on the DS-7001.
·
Never remove the CF card when the background color of the CF display is
red.
·
When the CF card status is “ERR.” or “FULL”, format the CF card.
·
When deleting the data on the CF card, format the CF card. The file can
not be completely deleted by the delete process on the PC.
·
Do not delete the REC folder.
·
Do not create other folder or file in the CF card.
·
When removing the CF card (media) from the PC, refer to the operation
manual of the PC, OS, card reader, and follow the correct procedure.
Sleep Function
·
When using the sleep function on the DS-7001, perform monitoring on the
central monitor. During the sleep mode, tone and lamp will not function,
and all alarm sound will not be generated on the DS-7001.
Maintenance
·
Clean the equipment frequently so stains can be removed easily.
·
To prevent injury, it is recommended to wear gloves when cleaning the
equipment.
·
Do not allow liquids or cleaning solution to enter the monitor or connectors.
·
Do not use organic solvents, thinner, toluene and benzene to avoid
damaging the resin case.
·
Do not polish the housing with abrasive or chemical cleaner.
·
When sterilizing the entire room using a spray solution, pay close attention
not to have liquids get into the monitor or connectors.
·
Use only neutral detergent to clean the housing. Do not use chemical
cloth, scrub brush, abrasive, polishing powder, hot water, volatile solvent
and chemicals (cleanser, thinner, benzine, benzol, and synthetic detergent
for house and furniture), or sharp-edged tools. The surface resin coating
may be damaged, resulting in discoloration, scratches, and other problems.
·
Do not open the housing of this device.
·
Avoid alcohol or other liquids from getting into the equipment.
·
The time-change components must be replaced at specified period.
Precautions for Use of SpO
2
Sensor
DANGER
Burn Risk in Using SpO
2
Sensor
In SpO
2
monitoring, always use the sensor/relay cable specified by
Fukuda Denshi. If any other sensor/relay cable is used, a high
temperature rise of the sensor may place the patient in danger of
burns.
If there are any questions regarding the sensor/relay cable use for
SpO
2
measurements of this device, please contact Fukuda Denshi
service representative.
Precautions for Use of NIBP Cuff
CAUTION
This product contains natural rubber latex which may cause allergic
reactions.
xv
Содержание DS-7001
Страница 28: ...xxvi Blank Page ...
Страница 40: ...1 6 Blank Page ...
Страница 46: ...2 6 Blank Page ...
Страница 57: ... SV 900 SV 300 Servo i Servo s CJ 501 Servo i Servo s CJ 502 SV 900 SV 300 CJ 500 3 11 ...
Страница 78: ...4 16 Blank Page ...
Страница 104: ...5 26 Blank Page ...
Страница 114: ...6 10 Reference For details of the battery lamp refer to 3 Preparation Before Use To Turn ON the Power ...
Страница 126: ...7 12 Blank Page ...
Страница 150: ...9 14 Blank Page ...
Страница 151: ...Printed in Japan 4R2160 200606 39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 ...