3-9
897N203029A FDC NX600 Instruction Manual 2022.01
3 Measurement
4. Load the dedicated sample rack for the
blood collection tube or FUJI sample
tube to be used.
NOTE
To avoid a suction error or malfunction,
use a sample tube that matches the la-
bel on the sample rack.
NOTE
Refer to Section 3.9 for sample tubes
that can be used.
Blood collection tube
or FUJI Sample Tube
5. Remove the cap of the blood collection
tube or FUJI Sample Tube that contains
sample, and then load it on the dedicat-
ed sample rack.
NOTE
Make sure there are no bubbles in the
sample in the blood collection tube or
FUJI Sample Tube. Use a centrifuge to
remove bubbles. If bubbles are present,
the sample may be detected incorrectly.
3