8 Troubleshooting
8-13
897N203029A FDC NX600 Instruction Manual 2022.01
(3) When there are insufficient number of or no mixing cups
Error
code
Error message
Error condition
Troubleshooting
W0151
Lack of mixing cup
Lack of mixing cups.
Set mixing cup and re-
run the tests from the
beginning.
There are not enough
mixing cups.
After you replace the mixing cups, re-
run from the measurement request.
W0152
No mixing cup
Set mixing cup and re-
run the tests from the
beginning.
E0154
Mixing cup detection
sensor ERR
Contact your dealer.
Error of the mixing cup
detection sensor.
If the error still occurs after switching
the analyzer off and on, please contact
your dealer for assistance.
(4) During a sample volume error
Error
code
Error message
Error condition
Troubleshooting
W0120
Sample shortage
Sample is insufficient.
Check the sample.
Either the sample is
not loaded or the sam-
ple volume is insuffi
-
cient.
Make sure that the sample volume in
the tube is within the measurable range
and an appropriate sample tube is used
in the sample rack, and then rerun from
the measurement request.
NOTE
If the sample volume is low, add
sample to the tube or perform
manual sample application.
W0060
Sample shortage
Check the sample vol-
ume.
Confirm that a sam
-
ple rack matching the
sample tubes is used.
There is not enough
sample volume.
W0140
Surface detection
ERR (sample)
Sample volume is too
much.
Check the sample vol-
ume.
There is too much
sample volume.
8
Содержание FUJI DRI-CHEM NX600
Страница 2: ......
Страница 33: ......
Страница 72: ......
Страница 106: ......
Страница 160: ......
Страница 167: ...8 Troubleshooting 8 3 897N203029A FDC NX600 Instruction Manual 2022 01 8...
Страница 248: ......
Страница 250: ......
Страница 254: ...11 USM User Safety Mode 11 4 897N203029A FDC NX600 Instruction Manual 2022 01 ADM All functions are operable 11...
Страница 266: ......
Страница 267: ......