![FujiFilm FUJI DRI-CHEM NX600 Скачать руководство пользователя страница 132](http://html1.mh-extra.com/html/fujifilm/fuji-dri-chem-nx600/fuji-dri-chem-nx600_instruction-manual_2343341132.webp)
6-16
897N203029A FDC NX600 Instruction Manual 2022.01
6 Maintenance
(3) Sample application unit
1. Make sure that the sample application
unit cover and two slide transfer weights
were removed for cleaning of the trans-
fer bar.
Sample application unit
2. Wipe the sample application unit with
gauze or a cotton swab moistened with
lukewarm or cold water.
(4) ISE unit probe
Probes
Make sure there is no dust on the six ISE probes.
When you clean the ISE probes, wipe off the dust
using a dry cotton swab.
NOTE
Do not use solvents, such as methyl alcohol
or isopropyl alcohol.
CAUTION:
ISE probes are needle-shaped parts. Do not touch
them directly by hand. Be careful not to bend the
probes.
6
Содержание FUJI DRI-CHEM NX600
Страница 2: ......
Страница 33: ......
Страница 72: ......
Страница 106: ......
Страница 160: ......
Страница 167: ...8 Troubleshooting 8 3 897N203029A FDC NX600 Instruction Manual 2022 01 8...
Страница 248: ......
Страница 250: ......
Страница 254: ...11 USM User Safety Mode 11 4 897N203029A FDC NX600 Instruction Manual 2022 01 ADM All functions are operable 11...
Страница 266: ......
Страница 267: ......