background image

2-3

897N203029A FDC NX600 Instruction Manual 2022.01

2 Part Names and Consumables

Light source lamp

Disposal box

RE

Reference fluid holder

Slide cartridge

Sample application 

unit cover

Sample rack

Diluent holder

Tip rack

Mixing cup

2

Содержание FUJI DRI-CHEM NX600

Страница 1: ...NX600 FUJI DRI CHEM NX600i 2nd Edition 897N203029A January 2022 INSTRUCTION MANUAL FUJI DRI CHEM CLINICAL CHEMISTRY ANALYZER 13 EN This Manual describes details on how to operate the FUJI DRI CHEM NX6...

Страница 2: ......

Страница 3: ...UJIFILM products due to the use of products of other manufacturers not supplied by FUJIFILM 5 FUJIFILM shall not be liable for malfunctions and damage resulting from remodeling maintenance and repair...

Страница 4: ...ith all the series FUJI DRI CHEM NX600 and FUJI DRI CHEM NX600i the former includes the PF Plasma Filter unit and the latter does not So for the latter sections that describe the unit can be skipped N...

Страница 5: ...8 1 17 FUJIFILM PF Plasma Filters 1 8 1 18 FUJI DRI CHEM AUTO TIPS FUJI Sample Tubes and FUJI DRI CHEM MIXING CUPS 1 8 1 19 Light Source Lamp 1 9 1 20 Recording Paper 1 9 1 21 Labels 1 10 1 22 Symbols...

Страница 6: ...s 5 2 5 3 Loading Mixing Cups 5 5 5 4 Loading Diluent 5 7 5 5 Loading Reference Fluid 5 9 6 Maintenance 6 1 Periodic Maintenance Items 6 2 6 1 1 Items for Daily and Periodic Inspection by the User 6 2...

Страница 7: ...urement Errors 8 33 8 2 12 PF Errors 8 36 8 2 13 Errors Related to Data Communication Sample Barcode Reader 8 37 8 2 14 Calibration Errors 8 39 8 2 15 Other Errors 8 42 8 3 Slide Code Table 8 44 9 Fun...

Страница 8: ...9 39 9 3 22 Password Setting 9 39 10 Other Functions 10 1 Data Communication 10 1 11 USM User Safety Mode 11 1 Operator ID Input When Measurement Start and In Mode Function 11 1 11 2 Difference of Op...

Страница 9: ...uiring special attention instructions that must be followed supplementary explanations etc 1 2 Operation Precautions Before using this equipment please read this Instruction Manual carefully so that y...

Страница 10: ...or because it cannot be disposed of as a general waste Not doing so may cause environmental pollution Before discarding the main unit you must delete the measurement results personal information store...

Страница 11: ...ic environment should be evaluated prior to operation of the equipment Do not use this equipment in close proximity to sources of strong electromagnetic radiation e g unshielded intentional RF sources...

Страница 12: ...aces near sources of heat such as heaters Places where the temperature may drastically change Places where the equipment is subject to vibration or its support table is unstable 1 Install the equipmen...

Страница 13: ...he load on each portion is approx 7kg Make sure that the load bearing of the table at the installation location is sufficient Failure to observe these precautions may result in unexpected errors or eq...

Страница 14: ...strial waste Process this waste in compliance with any applicable regulations in your country such as by incineration melting sterilization or disinfection Further more when entrusting disposal to ano...

Страница 15: ...ad QC card in the box it came in until the accompanying lot of slides has been used up NOTE When QC information exceeds 5 lots the oldest QC information is deleted Slides with out QC information canno...

Страница 16: ...Use the reference fluid included with the ISE slides for ISE measurements Refer to the Instructions for Use of the slides Use only specified products Otherwise errors or malfunctions may occur 1 17 FU...

Страница 17: ...bes blood collection tubes and FUJI DRI CHEM MIXING CUPS for each sample Do not reuse old ones Use only specified products Otherwise errors or malfunctions may occur When the power is shut off during...

Страница 18: ...ohazard label Power button High tempera ture caution label Lamp caution label Power switch Depress to power ON O Depress to power OFF Name plate NOTE The example above uses an illustration for the FDC...

Страница 19: ...n Description Warning Caution Important consult documents Biological risks High temperature caution Power button Power switch LAN terminal RS 232C connector USB connector Fragile item handle carefully...

Страница 20: ...ing Precautions Sign Description Date of manufacture EC REP Authorized representative in the European Community In vitro diagnostic medical devices CE mark of conformity Importer Consult an electronic...

Страница 21: ...The following is an illustration of the FDC NX600 full option model FDC NX600i has slight difference Printer Card reader LCD touch panel Power button Disposal box Sampler cover Front bottom cover Sam...

Страница 22: ...connector Power switch Power inlet Air filter Louver RS 232C connector USB connector LAN terminal For sample barcode reader only Do not connect other USB devices CAUTION Do not remove the cover to loo...

Страница 23: ...DC NX600 Instruction Manual 2022 01 2 Part Names and Consumables Light source lamp Disposal box RE Reference fluid holder Slide cartridge Sample application unit cover Sample rack Diluent holder Tip r...

Страница 24: ...2 4 897N203029A FDC NX600 Instruction Manual 2022 01 2 Part Names and Consumables Details of PF unit RE PF unit PF sensor suction pad PF reference plate PF sample rack PF pump 2...

Страница 25: ...Clear the entered information 5 Plasma filter PF production num ber lot display Displays the lot number of the plasma filter PF in use Make sure this number matches the production number lot of the pl...

Страница 26: ...n Manual 2022 01 2 Part Names and Consumables 2 2 2 Help Screen You can tap the button on the Measurement preparation screen to check the procedures about measurement maintenance consum able exchange...

Страница 27: ...2 7 897N203029A FDC NX600 Instruction Manual 2022 01 2 Part Names and Consumables 2 1 1 button The Measurement preparation screen ap pears 2 button Return to the previous screen 2...

Страница 28: ...aracters and symbols 1 2 5 4 7 6 3 1 Alphabetical characters numbers and symbols buttons Enter alphabetical characters numbers and symbols 2 button Delete all entered characters and then quit keyboard...

Страница 29: ...is selected When the button is selected When the ABC button is selected a 0 9 8 8 ABC button Display the entry screen for uppercase alpha betical characters and numbers 9 abc button Display the entry...

Страница 30: ...ion manual 1 Installation method 1 NOTE An AC power cable is not supplied as a bundled item Please prepare an AC power cable conforming to the specifications below The requirements of the cable specif...

Страница 31: ...FUJI PLAIN TUBE 1 5 mL 500 1 box FUJI PLAIN TUBE 0 5 mL 500 1 box Recording paper 6 rolls 1 box Light source lamp 1 O ring 4 Plasma Filter PF 50 1 bag PF Suction Pad 1 PF Tube 1 FUJI DRI CHEM MIXING C...

Страница 32: ...ght 1 Sample barcode reader 1 CAUTION The sample barcode reader specified for the FDC NX600 can be used Do not connect a barcode reader other than specified for the FDC NX600 Otherwise physical damage...

Страница 33: ......

Страница 34: ...Chapter 5 Refer to Section 3 2 1 Cleaning the disposal box Checking the recording paper Refer to Section 3 2 2 Starting the FDC NX600 Refer to Section 3 2 3 Checking the date and time Refer to Sectio...

Страница 35: ...Section 3 7 Basic measurements To test CRP and CM Colorimetric slide CRP CM Slide Slide set order Refer to Section 4 5 To test CM and ISE slide ISE Slide CM Slide To test CRP CM and ISE slide CRP CM S...

Страница 36: ...ox If the analyzer is used without the box the measurement results may be negatively affected WARNING Do not touch used slides and tips with your bare hands as this may cause infection If you acciden...

Страница 37: ...cover 2 Recording paper Check the remaining recording paper FUJI thermal paper If the recording paper is short for the meas urement replace the old roll with a new one Refer to Section 6 4 Cutter open...

Страница 38: ...been shut off during measurement due to a power outage or other issue replenish the tips refer to Section 5 2 and replace the mix ing cups refer to Section 5 3 The analyzer may lose the information o...

Страница 39: ...s and calibration and measurement results might be incorrect 5 Loading consumables Load the consumables tips mixing cups diluent and reference fluid in accordance with the type of measurement Refer to...

Страница 40: ...dical assistance if necessary 1 Checking the LCD When Ready to test appears on the LCD make sure that the analyzer is not applying samples NOTE You can stop sample application by press ing the STOP bu...

Страница 41: ...d allow them to reach room tempera ture for 5 minutes or more with the condition of the individual package before unwrapping and using them Otherwise the measurement results and the analyzer operation...

Страница 42: ...the sample rack NOTE Refer to Section 3 9 for sample tubes that can be used Blood collection tube or FUJI Sample Tube 5 Remove the cap of the blood collection tube or FUJI Sample Tube that contains s...

Страница 43: ...6 Select a sample type from the list 7 Select a dilution factor from the list Refer to Section 3 6 and Section 9 3 14 NOTE When the analyzer is connected to DMS sample information can be acquired aut...

Страница 44: ...h the sampler cover lock key WARNING Ensure the sampler cover is locked during sample processing to prevent injuries and infection When the display reads Ready to test the sam pler cover can be opened...

Страница 45: ...uction Manual 2022 01 3 Measurement Time until the end of measurement Measurement start time When the spotting operation is completed the screen on the left is displayed and the time until the measure...

Страница 46: ...ge To retest Section 4 4 tap the retest button NOTE For the details of the measurement results screen refer to Section 3 4 7 Ending measurement To complete the measurement unlock the sampler cover and...

Страница 47: ...easurement start scheduling for measurements using PF Refer to Section 4 3 After inversion whole blood samples start to get separated PF filtering may not occur normal ly and an error may occur NOTE T...

Страница 48: ...urement results that include values out side the reference interval are indicated by and the corresponding test name and value become red If tapped the details screen ap pears And will be indicated wh...

Страница 49: ...amount 4 Print button Print all measurement results of the dis played sample 5 Send button Send the measurement results of the dis played sample to a PC via DMS 6 Retest button Perform the measuremen...

Страница 50: ...play 1 2 1 Measurement month scroll buttons Change the display of the calendar month 2 Date buttons Display the measurement results screen of a date The currently displayed date is dis played in blue...

Страница 51: ...without adjustment NOTE Refer to the Instructions for Use of the slide for usable diluent for urine measurements NOTE Two tips are necessary for each slide one for sample the other for reference flui...

Страница 52: ...point ing right For details of the loading method see Section 3 3 3 NOTE The direction of the arrow on the slide is opposite to the slide transport direction NOTE When you measure an ISE slide and CM...

Страница 53: ...3 20 897N203029A FDC NX600 Instruction Manual 2022 01 3 Measurement 7 Tap the START button to start measure ment 3...

Страница 54: ...lly dilutes using the next well NOTE Do not dilute whole blood samples NOTE The default dilution factors designated by the manufacturer are as follows ISE urine tests 2x CRP tests 21x Excluding ISE te...

Страница 55: ...ables tips and mixing cups If consumables are insufficient supplement or replace them Loading tips Refer to Section 5 2 Loading mixing cups Refer to Section 5 3 2 Load the diluent Loading diluent Refe...

Страница 56: ...Instruction Manual 2022 01 3 Measurement 5 In the list tap x2 Dilution is performed according to Dilution setting in Mode function for administra tor None No dilution NOTE If necessary enter or choose...

Страница 57: ...k it with the sampler cover lock key WARNING Ensure the sampler cover is locked during sample processing to prevent injuries and infection When the display reads Ready to test the sam pler cover can b...

Страница 58: ...he dilution factor at the sample in formation registration screen When 2x has been set at the sample information registration screen the dilution factor is 42x Diluent for CRP 1 Check the quantity of...

Страница 59: ...n and then specify 3x Refer to Section 3 6 NOTE If necessary enter or choose other sample information 5 Close the sample cover and then lock it with the sampler cover lock key WARNING Ensure the sampl...

Страница 60: ...analyzer Sampler cover 3 Close the sampler cover 4 Make sure that measurement and mode operation are not in progress CAUTION Do not switch off the power during measurement or mode operation Otherwise...

Страница 61: ...CAUTION Shutdown will be automatically performed after the process completed Make sure NOT to turn the power switch off CAUTION Were you not able to operate on the screen press the power button at fro...

Страница 62: ...water and then disinfect the area Seek medical assistance if necessary WARNING Used slides and tips are infectious industrial waste Process this waste in compliance with any applicable regulations in...

Страница 63: ...tive liquid surface range of blood collection tubes that can be used Outer diameter of 16 indicates 15 0 to 16 6 mm and 13 indicates 12 0 to 13 3 mm The following illustrations show the design suction...

Страница 64: ...t general blood collection tube contents and the inspection items that can be measured Heparin blood collection tubes Inspection items Applicable to all items For GLU and NH3 measure the samples as so...

Страница 65: ...tion Make sure that the sample volume is sufficient by referring to the illustration to the left The top sur face of the sample fluid must be within the shaded section of the tubes in the illustration...

Страница 66: ...included in the bundled items NOTE The number in parentheses is the number of parts included in the bundled items NOTE The spacer 9 is are not used for the FDC NX600i Note There are upper and lower s...

Страница 67: ...le is down Insert the 0 pin as shown b Sample rack for 13mm 100mm Affix the label After setting the 8 spacer insert the 3 tube holder Set the 8 spacer H 32mm hole is down 0 Pin c Sample rack for 13mm...

Страница 68: ...e holder Affix the label 0 Pin f Sample rack for 16mm 100mm when using plasma filter PF Affix the label 0 Pin 0 Pin g Sample rack for 13mm 100mm when using plasma filter PF Affix the label 0 Pin 0 Pin...

Страница 69: ...3 36 897N203029A FDC NX600 Instruction Manual 2022 01 3 Measurement i Sample rack for calibrator CP Insert the 6 calibrator CP holder Affix the label 0 Pin 0 Pin 3...

Страница 70: ...varia tions in X of each lot 2 Inputting lot compensation coefficients using a QC card The lot compensation coefficients are recorded on QC cards and one QC card is packed with each box of slides Whe...

Страница 71: ...r lot are displayed on the LCD and output to the printer as well NOTE Do not insert another QC card while the slide test and lot number are displayed NOTE When the analyzer beeps and the test is not d...

Страница 72: ......

Страница 73: ...nfect the area Seek medical assistance if necessary IMPORTANT Perform the calibration measurement using Unit A mg dL IMPORTANT When the calibrator cap is opened there may be a liquid film on the bottl...

Страница 74: ...r concen tration settings CP1 CP2 and CP3 IMPORTANT When the calibrator concentration values are in correct calibration is not performed normally NOTE Refer to the Instructions for Use of the calibrat...

Страница 75: ...ali brators is displayed 6 Check the quantity of the consumables tips and mixing cups If consumables are insufficient tap the Set con sumables button to replenish or replace them Setting of tips Refer...

Страница 76: ...e slide cartridge and set them on the analyzer NOTE Set the slide cartridge securely so that it does not rise up Otherwise the slide reading error may occur c Open the lid of the CP and place it in th...

Страница 77: ...able Usage Methods 9 Tap the START button The CP1 calibra tion spotting screen is displayed 10 After CP1 sampling open the sampler cover take CP1 out from the sample rack and then firmly close the lid...

Страница 78: ...ibrator CP2 Perform b to d of Step 8 and then tap the START button The CP2 calibration spotting screen is dis played NOTE If a slide with a different lot number has been loaded an error screen is dis...

Страница 79: ...ling open the sampler cover take CP2 out from the sample rack and then firmly close the lid The CP3 calibration start measurement screen is displayed 13 Perform measurement of calibrator CP3 Perform b...

Страница 80: ...lose the lid The calibration measuring screen is displayed 15 When measurement ends the Calibra tion measurement result screen is dis played NOTE Print the calibration measurement re sults and store t...

Страница 81: ...CRP are displayed in a list Items are displayed in the order the QC card was read Items within their validity term have their dates displayed in black The following items are displayed in red the dat...

Страница 82: ...the concentration in the calibrator concentration display area at the Calibration Lot No screen are different 1 Tap the Settings button The administrator password entry screen ap pears 2 Enter the adm...

Страница 83: ...Applicable Usage Methods 3 Tap the Input field of each CP The software keyboard screen appears 4 Enter a value and then tap the OK button For the details of the software keyboard screen refer to Sect...

Страница 84: ...sage Methods 5 If the button is tapped the entered values are registered and the analyzer returns to the Calibration Lot No screen NOTE If the Reset button is tapped the en tered values are reset If t...

Страница 85: ...disinfect the area Seek medical assistance if necessary IMPORTANT In the case of a CM slide use a dedicated pipette that matches the sample application volume specifications in the Instructions for U...

Страница 86: ...so display the measurement results screen from the screen on the left and request a retest Refer to Section 4 4 5 Load slides For details of the loading method see Section 3 3 3 RE Manual pipette cove...

Страница 87: ...d Guide the tip using your side corner 8 Prepare sample pipetting Hold the pipette and touch the tip mounted on the pipette to the guide your side corner of the Manual pipette cover This helps positio...

Страница 88: ...is displayed Otherwise the measurement results will be inaccurate IMPORTANT When applying sample to a CM slide apply the sam ple within 1 mm of the center of the slide Otherwise the measurement resul...

Страница 89: ...MPORTANT Mount the tip securely into the nozzle of the pipette If mounting is poor the tip may be dislodged and the sample may scatter Applying samples to slides 0057 Make a drop and then move it near...

Страница 90: ...samples NOTE Use whole blood samples that have a hematocrit value Hct of 20 to 55 Up to about 350 L of plasma including 85 L of dead volume can be filtered The amount of filtered plasma decreases as...

Страница 91: ...F The PF card contains information to compensate for these differences 2 Using PF cards One PF card is included in one PF box with 50 filters Read a PF card in the same way a QC card is read Be sure t...

Страница 92: ...epare the consumables Prepare the tips mixing cups diluent and refer ence fluid as required Loading tips Refer to Section 5 2 Loading mixing cups Refer to Section 5 3 Loading diluent Refer to Section...

Страница 93: ...he sample volume is too high the sample may overflow from the blood collection tube when the PF is loaded There must be empty space in the tube Refer to Section 4 3 2 2 Mix the whole blood sample in t...

Страница 94: ...PF and then remove the sample container WARNING When discarding used PF always follow biohazard procedures wear gloves a lab coat and safe ty goggles in accordance with the sample handling rules of yo...

Страница 95: ...se blood collection tubes that may cause the tip of the sample suction nozzle of the PF to be physically blocked by additives or other materials NOTE As shown in the illustration below the required sa...

Страница 96: ...iously filtered by the PF IMPORTANT The retest button is designed only to automatically enter the sample number and sample ID of the previously measured sample Do not use this button to retest samples...

Страница 97: ...es and sample Refer to Section 3 3 NOTE If the amount of remaining sample is insufficient run the test with manual pi petting Refer to Section 4 2 2 Tap the retest button on the meas urement results s...

Страница 98: ...ck it with the sampler cover lock key WARNING Ensure the sampler cover is locked during sample processing to prevent injuries and infection When the display reads Ready to test the sam pler cover can...

Страница 99: ...RP CM and ISE slides CRP 4 5 2 4 ISE CRP CM slide Stack CM slides before CRP and ISE slides 4 Measuring CM slides that require dilution 5 2 4 5 2 2 5x dilution 2x dilution If slides are configured wit...

Страница 100: ...es to be measured from DMS The Work list index screen appears NOTE If there are no measurement requests the buzzer sounds and the message There is no request information is dis played NOTE The display...

Страница 101: ...of the regis tered sample information are displayed NOTE If the button is tapped registration is cleared NOTE The work list measurement items can be printed by tapping the button 3 Tap the OK button...

Страница 102: ...mber tap the No button at the measurement results screen Refer to Section 3 4 2 Enter the administrator password and then tap the OK button The Sample number entry screen appears 3 Edit the sample num...

Страница 103: ...e barcode reader before using it A maximum of 13 alphanumerical characters can be input for a sample ID Always use the sample barcode reader under lighting at 3 000 lux or lower and avoid direct sunli...

Страница 104: ...is complete the analyzer makes a beep sound The sample ID is displayed on the LCD and is registered NOTE Position the sample barcode reader properly facing to the barcode label Otherwise an incorrect...

Страница 105: ...using the sample barcode reader After reading is complete the analyzer makes a beep sound The operator ID is displayed on the LCD and is registered NOTE Position the sample barcode reader properly fac...

Страница 106: ......

Страница 107: ...ixing cups tap the display of consumables to display the loading screen When loading is complete press the Finish button to exit NOTE If there are not enough consumables tips mixing cups diluent refer...

Страница 108: ...s WARNING During sample application tap the STOP button to stop application and then load the tips Load ing is unsafe during sample application because the sampler and the consumable disk are mov ing...

Страница 109: ...Remove the tip rack from the consum able disk NOTE Always remove the tip rack from the consumable disk before loading tips 5 Load tips in all holes of the tip rack NOTE Always set tips in all holes o...

Страница 110: ...7N203029A FDC NX600 Instruction Manual 2022 01 5 Loading Consumables 7 Check that tips have been set in all holes of the tip rack And then tap the Replenish button of tips 8 Tap the Finish button to e...

Страница 111: ...nt uses one mixing cup If no mixing cups remain a measurement cannot be run WARNING During sample application tap the STOP button to stop application and then load the mixing cups Loading is unsafe du...

Страница 112: ...con sumable disk and then set new mixing cups WARNING Used mixing cups are infectious industrial waste Process this waste in compliance with any appli cable regulations in your country such as by inc...

Страница 113: ...ube for the diluent Otherwise measurement errors or malfunctions may occur The specified tubes for diluent are the FUJI PLAIN TUBE 1 5 ml and 0 5 ml IMPORTANT The diluent for CRP measurements evaporat...

Страница 114: ...ayed on the LCD make sure that the analyzer is not applying samples RE Diluent 3 Load a tube containing diluent NOTE Tap the consumables display on the top screen and select the Diluent tab to dis pla...

Страница 115: ...nsafe during sample application because the sampler and the consumable disk are moving IMPORTANT Use a FUJI PLAIN TUBE 0 5 or 1 5 mL for reference fluid Otherwise measurement errors or malfunctions ma...

Страница 116: ...ANT Clean the cap of the reference fluid every time you load the reference fluid Refer to Section 6 8 RE Reference fluid NOTE Make sure the reference fluid cap is closed properly NOTE Tap the consumab...

Страница 117: ...ntaining the analyzer get training from our technical support staff CAUTION Refer to each section for information about usable solvents for cleaning and decontamination CAUTION Do not use alcohol to c...

Страница 118: ...unit Once every 3 months or as appropriate such as when inaccurate measurement values occur Refer to Section 6 3 Recording paper Appearance of red lines on both sides of paper Refer to Section 6 4 Lig...

Страница 119: ...t interval Interference filter Once every 4 years Syringe Once every 3 years Incubator pressure plate Inspection once every 3 years WARNING After inspection always connect the power cord switch on the...

Страница 120: ...ORTANT If the air filters are not cleaned an increase in the temperature inside the analyzer may negatively affect the measurement values NOTE Clean all 4 filters 1 Switch off the analyzer power Louve...

Страница 121: ...water NOTE If you wash the filter with running water make sure it is dried well before setting in the analyzer 1 2 4 Insert the filter into the louver and put the louver back in the analyzer Push up t...

Страница 122: ...a slide transfer error or a tip ejecting error occurred clean the sample application unit WARNING Always wear gloves a lab coat and safety glasses while working If you accidentally touch a contaminant...

Страница 123: ...6 7 897N203029A FDC NX600 Instruction Manual 2022 01 6 Maintenance 6 3 1 Reference Plate Level Check Perform Reference plate level check before cleaning 1 Tap the button 2 Tap the Normal mode button 6...

Страница 124: ...pears 4 Check the results of Reference plate level check When has appeared in the check results cleaning is necessary Tap the OK button Using the procedure on the following pages clean the transfer ba...

Страница 125: ...e Transfer Bar Incubator Sample Application Unit and ISE Unit 1 Cleaning the transfer bar 1 Tap the Cleaning mode button in Normal mode 2 Remove the slide cartridge and the tip rack from the analyzer...

Страница 126: ...move the two slide transfer weights NOTE There are 2 types of slide transfer weight one each for CM tests and ISE tests The slide transfer weight for CM tests has hooks on the left so lift up from the...

Страница 127: ...application unit 9 Next clean the incubator sample appli cation unit and ISE test unit NOTE To quit the cleaning attach the removed slide transfer weight and sample ap plication unit cover at their or...

Страница 128: ...g the power cord before the following procedure 2 Incubator Panel unit pedestal 1 Open the sampler cover and panel unit Remove the panel unit pedestal Connector 2 Unplug the connector of the incubator...

Страница 129: ...kewarm water IMPORTANT Wring the cloth when it is soaked too much If wa ter gets inside the machine it will cause rust 6 Place the removed incubator on a flat surface with the pressure plates facing u...

Страница 130: ...ach pressure plate NOTE Wipe gently to avoid scratching as it will affect the measurement values IMPORTANT When wiping change the cloth surface for each cell Otherwise dirt may be transferred IMPORTAN...

Страница 131: ...white plate Reference black plate 12 Turn over the reference plate and clean the white plate and black plate with a dry cotton swab Photometer 13 Clean the photometer head with a dry cotton swab Thum...

Страница 132: ...ion unit 2 Wipe the sample application unit with gauze or a cotton swab moistened with lukewarm or cold water 4 ISE unit probe Probes Make sure there is no dust on the six ISE probes When you clean th...

Страница 133: ...the slide transfer weights 2 Load the sample application unit cover and tighten the thumbscrews 3 Load the incubator cover 4 While aligned with the groove plug the connector of the incubator cable in...

Страница 134: ...ording paper NOTE If the paper replacement was performed during a measurement the measure ment results might be printed out NOTE Set the paper roll so the paper is pulled from under the paper roll as...

Страница 135: ...0 hours CAUTION The light source lamp gets very hot Before cleaning the light source unit or replacing the light source lamp turn off the power wait at least 5 minutes and then check the lamp has cool...

Страница 136: ...cure so that metal hooks hold from above the lamp housing 5 Turn the thumbscrew of the lamp unit to the left remove the lamp and then re place it with a new one IMPORTANT Tighten the thumbscrews secur...

Страница 137: ...ew securely Otherwise the measurement results may be negatively affected 8 Attach the front bottom cover load the disposal box and then switch on the power of the FDC NX600 NOTE Put the front bottom c...

Страница 138: ...nant immediately rinse the contaminated area thoroughly un der running water and then disinfect the area Seek medical assistance if necessary IMPORTANT If you do not inspect and replace the sampler O...

Страница 139: ...e the mois ture from the O ring portion Remove any foreign objects remaining on the O ring 6 Perform a measurement and confirm that no errors occur IMPORTANT Do not store the AUTO TIPS with the storag...

Страница 140: ...6 24 897N203029A FDC NX600 Instruction Manual 2022 01 6 Maintenance 6 6 2 Inspecting the Sampler O ring 1 Tap the button 2 Tap the Normal mode button 3 Tap Sampler leak check 6...

Страница 141: ...e checked to select it and then run the leak check Once check ends the Check results screen ap pears CAUTION Keep your fingers clear as the sampler unit and consumable disk move 6 Check the results of...

Страница 142: ...re NG tap the Replace O ring button and then replace the O ring according to the following procedure NOTE The check results can be printed by tapping the button 7 Once the check ends open the sampler...

Страница 143: ...RTANT If you do not replace the O ring periodically the sample application volume will be inaccurate which may negatively affect the measurement values IMPORTANT Be sure to use the O ring designed for...

Страница 144: ...ap Replace O ring 5 Tap the START button The sampler nozzles turn to the replacement position CAUTION Keep your fingers away as the sampler unit moves 6 Follow the on screen instructions to replace th...

Страница 145: ...6 29 897N203029A FDC NX600 Instruction Manual 2022 01 6 Maintenance 7 Tap Leak check to check for leaks Refer to Section 6 6 2 6...

Страница 146: ...g If you accidentally touch a contaminant immediately rinse the contaminated area thoroughly un der running water and then disinfect the area Seek medical assistance if necessary WARNING Dispose of co...

Страница 147: ...rence fluid will be concen trated which may negatively affect the measurement values 1 If reference fluid is loaded remove the reference fluid tube and discard it Reference fluid cap packing 2 Remove...

Страница 148: ...DC NX600 Instruction Manual 2022 01 6 Maintenance Reference fluid cap packing 4 Set the reference fluid cap packing prop erly NOTE Do not touch the bottom side reference fluid side of the reference fl...

Страница 149: ...touch a contaminant immediately rinse the contaminated area thoroughly un der running water and then disinfect the area Seek medical assistance if necessary WARNING Dispose of contaminants used for cl...

Страница 150: ...C NX600 Instruction Manual 2022 01 6 Maintenance 3 Tap PF maintenance 4 Tap the Cleaning button 5 Remove the sample rack mixing cup and tip rack 6 Tap the OK button CAUTION Keep your fingers clear as...

Страница 151: ...according to the on screen in structions 8 Tap the OK button to end cleaning Surface of PF sensor glass surface PF Suction Pad black PF reference plate 9 Run the leak check Refer to Section 6 9 2 NOT...

Страница 152: ...tion Manual 2022 01 6 Maintenance 6 9 2 Inspecting the PF Suction Pad The PF suction pad must be inspected monthly and replaced every two years 1 Touch the PF Maintenance button as in the previous sec...

Страница 153: ...clean or replace the pad NOTE The check results can be printed by tap ping the button If the check results are NG tap the Cleaning button clean the PF unit and then run the leak test again Refer to Se...

Страница 154: ...Pad Replace a PF suction pad once every 2 years WARNING Process used PF Suction Pad as infectious waste in compliance with any applicable regulations in your country 1 Touch the PF Maintenance button...

Страница 155: ...6 39 897N203029A FDC NX600 Instruction Manual 2022 01 6 Maintenance RE Tip rack Mixing cups Sample rack 3 Remove the sample rack mixing cups and tip rack 4 Tap the button to switch off the power 6...

Страница 156: ...direction shown and remove it while loosening 6 Install the new PF Suction Pad by turn ing it in the opposite direction NOTE Be sure to tighten the PF Suction Pad securely 7 Switch on the analyzer 8...

Страница 157: ...r and unplug the power cord Replace the PF pump tube once every 2 years Replacement is also necessary if the PF has not been used for one month or longer Front bottom cover 1 Remove the disposal box a...

Страница 158: ...ad the new tube H on to the rotor NOTE As shown in the diagram to the left align the base of the joint E with the cover hole F NOTE Set so that the two front marks face the front Front Marks Cover hol...

Страница 159: ...aintenance 8 Attach the front bottom cover NOTE Put the front bottom cover back horizon tally as it was 9 Attach the disposal box 10 After the replacement run the leak check If the check results are N...

Страница 160: ......

Страница 161: ...slide and then meas ure 7 3 Monitoring Results Configure the target values and acceptable ranges for measurements from control fluid results and monitor for changes over time 7 4 Control Mode You can...

Страница 162: ...ntrol 7 2 897N203029A FDC NX600 Instruction Manual 2022 01 7 6 Control Chart 1 Tap the button 2 Tap the Control button 3 The measurement preparation screen of control mode appears 4 Tap the graph disp...

Страница 163: ...changed Control lot was changed Outside of accept able range 6 The control chart appears NOTE If the control liquid button is tapped you can change the control liquid to be dis played NOTE Measurement...

Страница 164: ...n outside of the acceptable range of control Acceptable range lower limit Acceptable range upper limit NOTE The acceptable range of control displays the value entered for Range setting for control Ref...

Страница 165: ...fore or during measurement or warnings appear in the measurement results the measurement results may be inaccurate Refer to the related troubleshoot ing pages and rerun the measurement 8 1 Error Displ...

Страница 166: ...ction Manual 2022 01 8 1 1 Error Log You can view the history of errors that occurred by using Error log in the function menu 1 Tap the button 2 Tap the Error log button An error log list is displayed...

Страница 167: ...8 Troubleshooting 8 3 897N203029A FDC NX600 Instruction Manual 2022 01 8...

Страница 168: ...diluent type error 8 2 10 8 32 E0200 E0203 E0210 E0213 E0220 E0227 Slide transfer error 8 2 6 8 21 E0300 E0301 E0302 Reference white plate measurement error during initial operation 8 2 8 8 24 E0509...

Страница 169: ...rror inside circuit board 8 2 8 8 25 E4200 E4201 Calibration error 8 2 14 3 4 8 39 E4202 E4203 Calibration error 8 2 14 5 8 40 E4204 Calibration error 8 2 14 6 8 41 E4205 Calibration error 8 2 14 5 8...

Страница 170: ...en read 8 2 9 3 8 29 W0140 Liquid volume error 8 2 4 4 8 13 W0141 W0144 Liquid volume error 8 2 4 7 8 16 W0151 W0152 Mixing cup error 8 2 4 3 8 13 W0170 QC card has not been read 8 2 9 2 8 29 W0173 Th...

Страница 171: ...13 1 8 37 W4511 Type 4 LAN communication error 8 2 13 1 8 37 W4520 Error related to work list 8 2 13 1 8 37 W4720 Clock time error 8 2 8 8 27 W4721 W4733 Signal system error inside circuit board 8 2 8...

Страница 172: ...he up per limit of the measurement range Refer to Section 8 2 11 OR Measurement value of the ISE test is lower than the lower limit of the measurement range Refer to Section 8 2 11 Temperature control...

Страница 173: ...analyzer will show for the ISE slides that expire 2023 or later or of lot number 130000 to 139999 NOTE Depending on the device settings there may be cases in which is not displayed and an expired sli...

Страница 174: ...ntact your dealer for assistance 2 White lines appear on printed characters or certain parts of characters are missing The printer head has malfunctioned Please contact your dealer for assistance 3 Pr...

Страница 175: ...a tip correctly and then rerun the test from the beginning Refer to Section 5 2 NOTE Set tips into all holes of the tip rack after this error has oc curred E0083 W0089 Reference fluid tip detection ER...

Страница 176: ...ondition Troubleshooting E0086 Sample tip eject ERR After analyzer shut down please check the disposal box Turn the power ON Tips cannot be dis carded After analyzer shutdown empty dis posal box and t...

Страница 177: ...ng the analyzer off and on please contact your dealer for assistance 4 During a sample volume error Error code Error message Error condition Troubleshooting W0120 Sample shortage Sample is insufficien...

Страница 178: ...em Please contact your dealer for assistance E0124 Aspiration clogging ERR diluted sample Clogging was detected during sampling Check the sample and diluent E0112 Spotting clogging ERR sample Clogging...

Страница 179: ...I CHEM AUTO TIPS are being used b Make sure that tips are not being reused c Make sure that there are no bub bles on the sample surface in the sample tube d Make sure that a sample rack that matches t...

Страница 180: ...o Section 5 5 NOTE Replace the diluent reference fluid container with a new one Do not add diluent or reference fluid W0142 Surface detection ERR reference fluid Reference fluid is too much Check the...

Страница 181: ...e the diluent and reference fluid and then rerun from the measurement request NOTE If the error occurs again there might be a piping system prob lem Please contact your dealer for assistance E0122 Spo...

Страница 182: ...lease contact your dealer for assistance E1000 E1001 E1003 W1004 Sampler up down motor ERR Problem moving the sampler up and down Perform inspection Contact your dealer Up and down move ment error of...

Страница 183: ...lean the reference plate Refer to Section 6 3 Clean the photometer Refer to Section 6 3 After cleaning switch on the power If the errors still occur after cleaning replace the light source lamp Refer...

Страница 184: ...minutes and then check the lamp has cooled 3 During an interference filter motor error Error code Error message Error condition Troubleshooting E0900 E0901 E0903 W0904 Filter motor ERR Problem with t...

Страница 185: ...ntact your dealer E0600 E0604 Incubator motor ERR Problem with the incu bator Perform inspection Contact your dealer Movement error of the incubator motor of the incubator E0700 E0701 E0703 W0704 Prob...

Страница 186: ...mperature b Switch off the power c Clean the air filter Refer to Section 6 2 d Make sure that thumbscrews of the incubator are tightened properly and the incubator cable is connect ed to the connector...

Страница 187: ...again NOTE After inserting the connector of the incubator cable you must turn it to the right to lock it Refer to Section 6 3 3 NOTE If the error occurs again please contact your dealer for assis tan...

Страница 188: ...r for assis tance E0510 Faulty A D controller Contact your dealer E0515 DA conversion ini tialization ERR Contact your dealer E0516 AD conversion reset ERR Contact your dealer E0517 AD conversion ini...

Страница 189: ...contact your dealer for assis tance E1750 E1755 E1757 Slide reading unit failure Contact your dealer If the error still occurs after switching the analyzer off and on please contact your dealer for as...

Страница 190: ...setting for reference intervals is correct W4465 W4466 W4468 Memory recovery WARNING Recovery performed from backup memory Check whether the QC card is read by con firming the lot number If not re re...

Страница 191: ...efault Contact your dealer in case the ERR repeats If the error occurs again please contact your dealer for assistance E4490 File deletion ERR Contact your dealer in case the ERR repeats Please contac...

Страница 192: ...he ERR repeats Failed to copy files from the USB memory USB memory read er ror has occurred Check the USB memory and retry ac cording to the error message If the error occurs again please contact your...

Страница 193: ...tion has not been read Error code Error message Error condition Troubleshooting W0170 No QC card info No QC information Read QC card The QC card of the slides loaded in the slide cartridge has not bee...

Страница 194: ...t faces the far left side If the error occurs re peatedly contact your dealer QC card read error Make sure the card is inserted at cor rect orientation W4011 QC card ERR Cannot save data Panel card re...

Страница 195: ...W0041 Incorrect sample type WARNING Although a PF sam ple rack is loaded the sample type setting is not P S The sample type set ting of PF measure ment does not match Set the sample type to P S W0042...

Страница 196: ...t name and rerun the measurement from the beginning W0175 No slide Slides have not been set Rerun the tests from the beginning No slides are loaded in the slide cartridge Load slides to be tested and...

Страница 197: ...ain ERR For ISE measurement turn the power SW OFF and ON Contact your dealer in case the ERR repeats Other parameters besides ISE can be measured An error was detected during a self test per formed be...

Страница 198: ...malfunction Drift error and impedance error Impedance error and outside of the measurement range Drift error and outside of the measurement range Drift error impedance error and outside of the measure...

Страница 199: ...started Check ISE measurement errors that oc curred before this error and then carry out the corresponding troubleshooting After troubleshooting turn the power of the device on and off and then start...

Страница 200: ...than PF measurements can be measured with out turning the power SW OFF and then ON A PF measurement error has occurred and PF measurement cannot be started Check PF measurement errors that oc curred...

Страница 201: ...on of the communication cable Contact your dealer in case the ERR repeats LAN communication error W4511 Type 4 LAN commu nication ERR Check whether the set ting for type 4 or host computer is correct...

Страница 202: ...on The sample barcode reader is not connected Connect the sample barcode reader to the USB connector and then turn the analyzer off and on W5331 Exceeded number of characters from bar code reader The...

Страница 203: ...t suited for calibration has been loaded during the cali bration process Check the slide Rerun the calibration process again W0191 Wrong slide lot for calibration A different lot of slide other than d...

Страница 204: ...2 Calibration curve set ting ERR span error Misplacement of cali brator Check the level and expiration date of the calibrator and perform the calibration from the beginning The wrong calibrator may ha...

Страница 205: ...ror message Error condition Troubleshooting E4204 Calibration measure ment ERR Error in measured result After coping with the printed error perform the calibration from the beginning Error marks such...

Страница 206: ...rogress The disposal box is full of slides and tips While a measurement is not being per formed discard the slides and tips in the disposal box NOTE When there are 150 slides or 75 tips in the disposa...

Страница 207: ...e interval range setting control fluid setting and corre lation coefficients have been reset Check the settings The settings have been reset Check settings for the reference in terval range and the co...

Страница 208: ...PT ALT P 32 50 TCHO P 14 50 CPK P 33 50 NH3 P 15 50 LDH P 34 50 TG P 16 50 ALP P 35 50 CRE P 17 50 LAP P 37 50 TP P 18 50 CKMB P 38 50 ALB P 20 50 CHE P 39 50 TBIL P 21 50 AMYL P 43 50 Ca P 23 50 LIP...

Страница 209: ...1 1 Control 9 1 START Manual pipetting 4 2 Normal mode Date Time settings 9 2 1 Brightness Volume Display brightness 9 2 2 Error sound Other general sound Lamp management Lamp ON time 9 2 3 Lamp off t...

Страница 210: ...atient name 9 3 8 Patient ID Patient name Correlation coefficients 9 3 9 Lot compensation coefficients 9 3 10 Standard curve coefficients 9 3 11 Lot Qualification 9 3 12 PF calibration coefficients 9...

Страница 211: ...ng Initial values disabled a 1 b 0 of the function menu Measurement units follow Unit settings of the function menu Dilution conditions dilution is not normally performed This can be configured at the...

Страница 212: ...keyboard Section 2 3 3 Select the control liquid from the list when nec essary 4 Select the control liquid lot from the list when necessary IMPORTANT When you configure the control liquid lot make sur...

Страница 213: ...ple cover and then lock it with the sampler cover lock key WARNING Ensure the sampler cover is locked during sample processing to prevent injuries and infection When the display reads Ready to test th...

Страница 214: ...ing control mode always tap the HOME button to clear control mode Display of measurement results Green when within the acceptable range of control Red when outside of the acceptable range of control A...

Страница 215: ...settings Reference section Date Time settings 9 2 1 Brightness Volume Display brightness 9 2 2 Error sound Other general sound Lamp management Lamp ON time 9 2 3 Lamp off timer setting Lot informatio...

Страница 216: ...power off will appear when is tapped to exit Normal mode After the shutdown reboot the analyzer 9 2 2 Brightness Volume You can configure the brightness of the LCD and buzzer sounds error warning meas...

Страница 217: ...the cumulative illumination time of the lamp after replacing the lamp If the Lamp off timer setting checkbox is select ed the lamp is switched off automatically after some time You can configure the...

Страница 218: ...ulated Parameter 9 3 3 Unit conversion 9 3 4 Switch measurement range display 9 3 5 Reference interval range setting 9 3 6 Sample No ID settings 9 3 7 Work list selection setting Sample ID Patient nam...

Страница 219: ...the setting to take effect The confirmation di alog for the power off will appear when is tapped to exit Mode function for adminis trator After the shutdown reboot the ana lyzer 9 3 2 Host Connection...

Страница 220: ...tput from the printer You can choose from 4 communication types Type 1 Bidirectional connection Type 2 Unidirectional connection bidirectional format Type 3 Unidirectional connection standard format T...

Страница 221: ...on Tap MLLP setting to set MLLP Minimal Lower Layer Protocol in communication Full MLLP is applied to full message in com munication Custom MLLP is applied to only at the end of message in communicati...

Страница 222: ...T GOT value GPT value AST ALT AST value ALT value NOTE The calculation test name can be changed by Test item name setting in the Mode function for administrator Refer to Section 9 3 20 Sodium Potassiu...

Страница 223: ...e function for administrator precautions for use of calculation items are displayed To use calcula tion items after confirming the precautions tap the OK button Configure the calculation items to be u...

Страница 224: ...U L kat L 0 0167 NH3 15 g dL mol L 0 7139 LDH 34 U L kat L 0 0167 TG 16 mg dL mmol L 0 01129 ALP 35 U L kat L 0 0167 CRE 17 mg dL mol L 88 4 LAP 37 U L kat L 0 0167 TP 18 g dL g L 10 CKMB 38 U L kat...

Страница 225: ...splay method for data outside the measurement range passing the upper lower limit of the determination range for all tests If is configured an symbol is added to indicate the displayed measurement val...

Страница 226: ...3 4 Furthermore you can set a reference inter val for each sample type P S W or U of a reference interval name Adding and editing of reference interval names Next page buttons Tap the button of the it...

Страница 227: ...01 9 Function Menu Changing the display order Next page buttons Tap the button of the test for which you want to change the display order Tap the field of the test for which you want to con figure the...

Страница 228: ...nfigure whether to display the sample number in the measurement results under Edit and print No Configure whether to record the sequence num ber of the analyzer under Store No 9 3 8 Work List Selectio...

Страница 229: ...document IMPORTANT Incorrect entry of a and b causes incorrect measurement results Always make sure that the entries a b are correct Next page buttons Tap the Data input button NOTE If the button is t...

Страница 230: ...value To change the sample type tap the P S W or U button NOTE If the Reset button is tapped a reset con firmation screen appears To reset the dis played correlation coefficients for the sam ple type...

Страница 231: ...alyzer memory the FDC NX600 performs compensation calculations internally us ing the formula X Y b a In this way measured data obtained using the FDC NX600 are adjusted to match measured data that wou...

Страница 232: ...bove the meas urement range If the regression formula is Y 0 98X 5 A 5 mg dl B 607 mg dl Y mg dl X mg dl FDC reference Y 1 1X 5 Y X 0 10 600 0 600 A B Measurement using your instruments If the regress...

Страница 233: ...er was changed even if the new values c d and e are entered by us ing this setting the new standard curve information will not be updated It is necessary to read the QC card when the slide type number...

Страница 234: ...oard refer to Section 2 3 appears so you can enter the value NOTE The most significant digit of a Lot number the type number is fixed Enter from the second digit NOTE Incorrect entries will cause inco...

Страница 235: ...results Always make sure that the entered coefficients are correct NOTE The calibration curve coefficients of this setting are entered automatically by QC card read ing and calibration Configuring th...

Страница 236: ...button is tapped the calibra tion curve coefficients are printed 4 First tap the Input field of Lot The Lot number entry screen software key board screen appears 5 Enter a lot number and then tap the...

Страница 237: ...d screen appears 2 Enter B through F using the same proce dure Configuring the expiration date for use For the expiration date for use EXP enter xxxx xx four digits for the year and two digits for the...

Страница 238: ...bled It is possible to dis qualify slide LOTs which are registered in analyzer 1 Select a slide name which needs to change the setting 2 Dis qualification for selected slide LOT can be changed by sele...

Страница 239: ...apped all PF calculation correction coefficients are printed NOTE If the Reset all tests button is tapped a reset confirmation screen appears To reset the PF calculation correction coefficients of all...

Страница 240: ...re keyboard refer to Section 2 3 appears so you can enter the value NOTE If the Reset button is tapped a reset con firmation screen appears To reset the PF calculation correction coefficients of selec...

Страница 241: ...rement item and sample type PS and U When a dilution factor is not configured for each sample on the sample information registration screen measurement is performed using the dilution factor config ur...

Страница 242: ...7N203029A FDC NX600 Instruction Manual 2022 01 9 Function Menu 3 For the dilution factor tap a button and then select from None 2 3 4 5 or 10 NOTE In the case of CRP the selection menu to the left app...

Страница 243: ...ed Result Configure the print quantity 1 to 3 sheets and print order of measurement results Tap the print quantity you want to configure For Print order you can select Slide stacking order or Fixing N...

Страница 244: ...ars so you can edit the value IMPORTANT If the entry of the acceptable range is incorrect the judgement of the control measurement results is incorrect Always make sure that the entered values are cor...

Страница 245: ...rement tap the checkbox to select it NOTE By default the checkbox is cleared not enabled 9 3 19 Operator ID Setting NOTE When the USM function is disabled the following setting will appear In case tha...

Страница 246: ...E If ALP LDH is selected ALPi LDHi ap pears the same as ALP LDH Check this setting again before testing 3 Choose the item names to be used Tap either the BUN or UREA button NOTE When the name of BUN i...

Страница 247: ...en tapping the Mode function for administrator button on the Function menu screen the pass word input may be required The password can be changed here If the input field is tapped the software keyboar...

Страница 248: ......

Страница 249: ...prepare the software and a cable to receive data Before you use this function for the first time please contact your dealer for assistance NOTE Use a shielded LAN cable Otherwise electromagnetic inte...

Страница 250: ......

Страница 251: ...nction menu button is tapped an Operator ID input display is shown Then input Operator ID If input operator ID is registered and also within expiry the meas urement starts or the function menu becomes...

Страница 252: ...rator ID level NOTE OPE Operator ADM Administrator EMG Emergency Measurement Result OPE ADM When expired slide is used EXP is indi cated instead of measurement result EXP is used not only on screen in...

Страница 253: ...SM User Safety Mode 11 3 897N203029A FDC NX600 Instruction Manual 2022 01 Mode function OPE EMG Operable function is limited to Control Calibration START Manual pipetting and some maintenance function...

Страница 254: ...11 USM User Safety Mode 11 4 897N203029A FDC NX600 Instruction Manual 2022 01 ADM All functions are operable 11...

Страница 255: ...le name dcn_uid_yyyymmddhhmmss dat Ex dcn_uid_20220425201612 dat For the detail see the manual of the Operator ID Editor 2 Create the following folder structure in your USB Rename the file which is cr...

Страница 256: ...1 6 897N203029A FDC NX600 Instruction Manual 2022 01 5 The left pop up is shown Then set USB and after a few seconds select OK 6 The left pop up will be shown when im port is complete Take off the USB...

Страница 257: ...1 7 897N203029A FDC NX600 Instruction Manual 2022 01 11 3 2 Operator ID Export 1 Select Operator ID setting in ADM mode 2 Select Operator ID export 3 The left pop up is shown Then set USB and after a...

Страница 258: ...be shown when ex port is complete Then tap OK and take off the USB 5 Operator ID data dcn_uid dat is stored in below folder structure below in the USB WData GUI dcn_uid dat Browsing and editing of dcn...

Страница 259: ...g Registered Operator ID Registered Operator ID can be edited as the following procedure NOTE Editable items are the expiry of ID and the level 1 Select Operator ID setting in Mode function for admini...

Страница 260: ...User Safety Mode 11 10 897N203029A FDC NX600 Instruction Manual 2022 01 4 To change the expiry touch the expiry field and edit it using the software key board 5 To change the level select either ADM...

Страница 261: ...ting Emergency ID Dis qualification of registered Emergency ID can be set Also measurement limit count by use of Emergency ID can be set 1 Select Operator ID setting in Mode function for administrator...

Страница 262: ...eset the current count tap Reset button To set a new limit count select either 0 50 or 100 In case of 0 emergency ID does not function NOTE Measurement count by use of Emergency ID cannot exceed the l...

Страница 263: ...urement accuracy Within 0 005 OD 600 nm Display 7 inch color touch panel Backup data of test results For 270 samples Printer Thermal type paper size 58 mm x 25 m Warming up time Approx 10 minutes 25 C...

Страница 264: ...FDC NX600 Instruction Manual 2022 01 Effective period of service service life 6 years after installation This period is valid as long as the precautions for use are followed and regular periodic main...

Страница 265: ...d after abbreviations on the display and printout messages Abbreviation Meaning AD Analog Digital ADM administrator ave average Bef comp Before compress cali calibration charac character comm communic...

Страница 266: ......

Страница 267: ......

Страница 268: ...26 30 NISHIAZABU 2 CHOME MINATO KU TOKYO 106 8620 JAPAN https www fujifilm com EU Heesenstrasse 31 40549 Duesseldorf Germany Oudenstaart 1 5047 TK Tilburg The Netherlands...

Отзывы: