4-13
897N203029A FDC NX600 Instruction Manual 2022.01
4 Applicable Usage Methods
4.2 Manual Spotting (Colorimetric Slide)
4.2.1 Measuring by Manual Spotting (Colorimetric Slide)
WARNING:
When handling samples (blood or urine), always follow biohazard procedures (wear gloves, a lab
coat, and safety goggles) in accordance with the sample handling rules of your facility. If you acci-
dentally touch a sample, immediately rinse the contaminated area thoroughly under running water,
and then disinfect the area. Seek medical assistance if necessary.
IMPORTANT:
In the case of a CM slide, use a dedicated pipette that matches the sample application volume
specifications in the "Instructions for Use" of the slide. Carefully read the operation method of the
pipette in Section 4.2.2. (Be aware that the sample application volume of GLU-W is 6 µL, which is
different from other slide items.)
NOTE
Manual sample application is not possible with ISE slides.
NOTE
When the [Dilution] button was used to configure the dilution factor in manual sample appli
-
cation mode, the measurement value is calculated by a dilution factor.
1. Prepare the pipette, slides, and a sample.
2. Tap the button.
4
Содержание FUJI DRI-CHEM NX600
Страница 2: ......
Страница 33: ......
Страница 72: ......
Страница 106: ......
Страница 160: ......
Страница 167: ...8 Troubleshooting 8 3 897N203029A FDC NX600 Instruction Manual 2022 01 8...
Страница 248: ......
Страница 250: ......
Страница 254: ...11 USM User Safety Mode 11 4 897N203029A FDC NX600 Instruction Manual 2022 01 ADM All functions are operable 11...
Страница 266: ......
Страница 267: ......