9-27
897N203029A FDC NX600 Instruction Manual 2022.01
9 Function Menu
9.3.11 Standard Curve Coefficients
The calibration curve coefficients (A, B, C, D, E, and F) of the QC card information can be entered.
IMPORTANT:
Incorrect entry of calibration curve coefficients causes incorrect measurement results.
Always make sure that the entered coefficients are correct.
NOTE
The calibration curve coefficients of this setting are entered automatically by QC card read
-
ing and calibration.
<Configuring the lot number>
1. Tap the button of the measurement item
for which the calibration curve is to be
edited.
NOTE
Only CRP items for which QC cards
have been read are displayed.
List display area
2. From the list display area, tap the lot
number for which the calibration curve
is to be edited.