9-29
897N203029A FDC NX600 Instruction Manual 2022.01
9 Function Menu
<Configuration of calibration curve coefficients (A, B, C, D, E, and F)>
1. Enter the calibration curve coefficient
(A:), and then tap the [OK] button.
The Calibration curve coefficient (B:) entry
screen (software keyboard screen) appears.
2. Enter B through F using the same proce-
dure.
<Configuring the expiration date for use>
For the expiration date for use (EXP), enter xxxx-
xx (four digits for the year and two digits for the
month).
NOTE
Also enter the hyphen (-).
IMPORTANT:
Incorrect entries will cause incorrect measurement results.
Always make sure that the displayed values are correct.
NOTE
When the
button is tapped, values of the calibration curve coefficient configuration re
-
sults are printed.