9 Function Menu
9.3.12 Lot Qualification
This function button appears (enabled) when USM function is enabled. It is possible to (dis-)qualify
slide LOTs which are registered in analyzer.
1. Select a slide name which needs to
change the setting.
2. (Dis-)qualification for selected slide LOT
can be changed by selecting [ON] or
[OFF] button.
NOTE
When the disqualified slide lot is meas
-
ured, the test result will be indicated as
“CHK QC” on screen, print and in com-
munication.