1 Safe Usage and Handling Precautions
1.7 Caution Regarding Moving Parts
Do not place your fingers near moving parts (sampler and consumable disk) when operating the
equipment, including during maintenance.
Also be sure not to get your fingers, hair, clothing, or accessories caught in moving parts.
During sample processing, ensure the sampler cover is closed and locked with the sampler cover
lock key to prevent injuries and biohazards. If your finger touches a tip that has sample inside, the
sample may scatter and cause infection.
When the display reads [Ready to test], the sampler cover can be opened.
1.8 Installation Locations
Plug the power cable of the equipment into an outlet with a grounding receptacle. Otherwise, there
is risk of electrical shock.
Avoid the following installation sites.
• Places where spills or water leakage may occur.
• Places where the equipment is exposed to direct sunlight.
• Places near sources of heat such as heaters.
• Places where the temperature may drastically change.
• Places where the equipment is subject to vibration or its support table is unstable.
(1) Install the equipment in the following environmental conditions.
Installation environment:
Indoor use (do not expose to direct sunlight)
Illumination:
6,000 cd/m
2
(lux) or less(3,000 cd/m
2
(lux) or less when using
the sample barcode reader)
Atmospheric pressure:
810 hPa or more (equivalent to altitude 2,000 m)
Transient overvoltage category: II
Pollution degree:
2
Ingress protection rating:
IP00 (No Protection)
Operating temperature:
15 to 32 °C
Operating humidity:
30 to 80% RH (no vapor condensation)
(2) Use the equipment under the following electrical specifications.
Power supply voltage:
single phase AC 100 - 240 V ±10%
Frequency:
50 to 60 Hz
Consumption current:
1.9 - 0.8 A
Type of protection against
electrical shock:
CLASS 1 EQUIPMENT
(3) Use an independent outlet separate from other equipment.
(4) Make at least 10 cm of space behind and beside the equipment.
(5) Unplug the equipment from the AC outlet if it will not be used for an extended period of time.
WARNING:
WARNING:
WARNING:
CAUTION:
1
Содержание FUJI DRI-CHEM NX600
Страница 2: ......
Страница 33: ......
Страница 72: ......
Страница 106: ......
Страница 160: ......
Страница 167: ...8 Troubleshooting 8 3 897N203029A FDC NX600 Instruction Manual 2022 01 8...
Страница 248: ......
Страница 250: ......
Страница 254: ...11 USM User Safety Mode 11 4 897N203029A FDC NX600 Instruction Manual 2022 01 ADM All functions are operable 11...
Страница 266: ......
Страница 267: ......