ZIPThaw™ 202 System User Manual
Fremon Scientific, Inc.
5726 La Jolla Blvd., Suite 304, La Jolla, CA 92037USA
P
age 25 of 43
FDA
CLEARED
ZipThaw
™
20 2 System
User Manual
FIGURE 29. ZipSleeve
TM
Orientation.
7.9.2 LOADING THE ZIPSLEEVE INTO THE CHAMBER
1.
Open the Chamber door and place the loaded ZipSleeve
TM
inside the Chamber.
2.
Align the 3 slots at the top of the ZipSleeve with the 3 hangers inside the Chamber.
Note:
Please note that the ZipSleeve
TM
can be loaded only in the orientation indicated by the slots in the hard
plastic strip at the top of the ZipSleeve
TM
. If the plasma bag was frozen with the ports end of the bag folded over,
the cycle time may be extended. For larger volume units, that were frozen folded, you can do a cycle, and stop
after several minutes, unload, straighten the ports, reload and do the complete cycle.
Optimal freeze
configuration is frozen flat, ports out.
FIGURE 30. Device Open Door
Note:
ZipSleeve
TM
Maximum
Number of Uses is 8