7.10 END OF THAWING CYCLE
Upon thawing cycle completion of Chamber A, the following screen will be displayed:
FIGURE 36. Thawing Cycle Chamber A Complete & Chamber B Loaded and Running
Open the Chamber door and remove the ZipSleeve
TM
from the Chamber and than take out the thawed plasma bag.
Note:
If product is left inside chamber, system starts to count the time post End-of-Thawing Cycle, will alert the user
(i.e. sound & LED light will flicker) once, and again for 1 second, after 5minutes . Operator
must remove
the thawed
product from ZipThaw202 as soon the cycle completes.
Once removed, immediately close the Chamber Door!
7.11 TIMEOUT THAWING CYCLE
The ZipThaw
TM
system has a timeout mechanism that stops the thawing cycle in cases where the thawing process
doesn’t reach the target temp (i.e. 3
1°C).
Note:
Currently, the timeout is set to 60 minutes for each thawing cycle.
System will display the following message:
FIGURE 37. TimeOut Screen
20°
18:00 min
B
System Error
OK
Thawing Exceeded Max. Time!
Please Remove the Bag From
Chamber B.
01:18 min.
•
The Bag icon turns
GREEN.
•
A One Time Alarm will be
sounded for 1 Sec.
•
Chamber LED turns
Green
31
˚