10 USERS MESSAGES
The ZIPThaw
TM
202 system will prompt the user with operational and error messages as follows:
PHASE
MESSAGE
DESCRIPTION
BUZZER
SYSTEM RESPONSE
ACTION NEEDED
POW
ER
UP
S
C
R
EEN
“
Chamber A
Door is Open.
Please Close
Chamber Door
”
Warning Message:
Chamber Door A is
open.
NO
A warning Pop-Up
message that the
Chamber Door is
Open.
1.
Close the door to
resume power-up
sequence.
2.
Pop-Up massage will
disappear OR when
pressing “
OK
”
.
“
Chamber B
Door is Open.
Please Close
Chamber Door
”
Warning Message:
Chamber Door B is
open.
NO
A Warning Pop-Up
message that the
Chamber Door is
Open.
1.
Close the door to
resume power-up
sequence.
2.
Pop-Up massage will
disappear OR when
pressing “
OK
”
.
“
System Error,
FMS Technical
Support
Required
“
Error Messages:
Error
found during system
BIT Testing.
YES
A System Error Pop-
Up message.
Operational Halt.
1.
Switch off the Main
power on the back of
the device.
2.
Unplug the Power
Cord for 30 seconds.
3.
Restart device.
“Chamber A
Error, Use
Chamber
B”
Error Messages:
Error
found during system
BIT Testing: Chamber
A malfunction.
YES
A System Error Pop-
Up message.
Chamber A will be
disabled.
1.
Pressing
“
OK
”
will
close the popup.
2.
Follow IFU regarding
Technical Support
assistance!
“Chamber B
Error, Use
Chamber A
”
Error Messages:
Error
found during system
BIT Testing: Chamber
B Malfunctioning.
YES
A System Error Pop-
Up message.
Chamber B will be
disabled and the
specific Chamber
GUI will be modified.
1.
Pressing
“
OK
”
will
close the popup.
2.
Follow IFU regarding
Technical Support
assistance!
SE
LE
C
T
C
H
A
M
BE
R
SC
R
EE
N
“
Select
Chamber
”
System Prompt:
Indicating User choice
to use the available
Chamber.
NO
None
Follow IFU
PHASE
MESSAGE
DESCRIPTION
BUZZER
SYSTEM RESPONSE
ACTION NEEDED
IN
SE
RT
DA
TA
SC
RE
EN
“
Enter Required
Fields
”
System Prompt:
User to enter required
thawing cycle data.
NO
A System Pop-Up
message.
Pressing
“
OK
”
will close
the popup.
Follow IFU.
“
Incorrect ID,
Please Re-Enter
ID
”
System Warning:
When an wronge ID
was typed.
NO
A System Pop-Up
message.
Pressing
“
OK
”
will close
the popup.