ZIPThaw™ 202 System User Manual
Fremon Scientific, Inc.
5726 La Jolla Blvd., Suite 304, La Jolla, CA 92037USA
P
age 35 of 43
FDA
CLEARED
ZipThaw
™
20 2 System
User Manual
THAW
IN
G
IN
PR
O
G
RE
SS
“
Chamber A
Failure,
Technical
Support
Required
”
Error Message:
Indicating that upon
thawing cycle
“START”, Chamber A
malefunctioned.
YES
1.
A System Pop-
Up a message.
2.
The system
disables
Chamber A,
after 3 test
trials.
3.
The System
allows Chamber
B to continue.
1.
Pressing “
OK
” will
close the popup
message.
2.
Follow IFU regarding
Technical Support
assistance!
“
Chamber B
Failure,
Technical
Support
Required
”
Error Message:
Indicating that upon
thawing cycle
“START”, Chamber A
malefunctioned.
YES
1.
A System Pop-
Up a message.
2.
The system
disables
Chamber B,
after 3 test
trials.
3.
The System
allows Chamber
A to continue.
1.
Pressing “
OK
” will
close the popup
message.
2.
Follow IFU regarding
Technical Support
assistance!
“
Chamber Door
A Opened
”
“
Resume
Thawing?
”
Warning Message:
Indicating that
Chamber A Door was
opened during
thawing cycle.
YES
1.
A System Pop-
Up a message.
2.
The system
pause thawing
at Chamber A.
1.
Close Chamber Door.
2.
Pressing “
YES
”
will
resume thawing.
3.
Pressing
“
NO
”
will
abort the thawing
cycle.
“
Chamber Door
B Opened
”
“
Resume
Thawing?
”
Warning Message:
Indicating that
Chamber B Door was
opened during
thawing cycle.
YES
1.
A System Pop-Up
a message.
2.
The system
pause thawing at
Chamber B.
1.
Close Chamber Door.
2.
Pressing “
YES
”
will
resume thawing.
3.
Pressing
“
NO
”
will
abort the thawing
cycle.
“
Thawing
Exceeded Max.
Time!
Please Remove
the Bag From
Chamber A/B
”
Error Message:
Indicating that the
system timeout the
thawing cycle due to
failure to thaw the
bag to the target
temperature within a
predefined-
timeouttime.
YES
1.
A System Pop-
Up a message.
3.
The system
terminate the
thawing process
at the said
chamber.
4.
Press “
OK
” and open
the chamber door
and remove the
ZipSleeveTM from
the chamber and
than take out the
thawed plasma bag.
PO
ST T
H
A
WI
N
G
CYCLE
“Chamber A
Door Opened,
Thawing Cycle
Stopped!”
Error Message:
Indicating that
Chamber Door A was
left open for more
than 5 minutes.
YES
1.
A System Pop-Up
a message.
The system
terminates thawing
at Chamber A.
1.
Pressing
“
OK
”
will
close the popup
message.
2.
Chamber A GUI will
revert to
“
Select
Chamber
”
screen.