The following error message will appear:
FIGURE 9. CHAMBER ERROR POP-UP SCREEN
Pressing
“
OK
”
will make the error message disappear and the following screen will be displayed where the faulty
Chamber will be disabled and will have a grayed-out graphic appearance.
FIGURE 10. HOME SCREEN WITH CHAMBER ERROR
7.3.3 SYSTEM RESET RECOVERY
In case that the system defaults to an unexpected “Reset” mode,
upon successful BIT , the home screen will
appear and the system will recover in to pre-thawing state see 7.7, Fig 24.
Note:
It is the user
’s
discresion to start a new thawing cycle, using these bags or discarding them.
To continue a new thawing cycle proceed to 7.3.2.
7.3.4 SUCCESSFUL POWER UP
Upon successful pre-heating cycle, the home screen will appear,.
The screen will prompt: “
Select Chamber
” as