Operating Manual CMGZ411/421
38
Scaling instrument A+B CMGZ421
Purpose:
With a setting of 1.00, the analogue output (0...10V) provides the nominal voltage at nominal
tension force. By decreasing the scale parameter, the nominal voltage is decreased; if the scale
parameter is increased, nominal voltage is also increased.
Range:
0.01 to
10.00
Default:
1.00
Increment:
0.01
Unit:
[
-
]
Scaling instrument A CMGZ421
Purpose:
With a setting of 1.00, the analogue output (0...5V) provides the nominal voltage at nominal
tension force. By decreasing the scale parameter, the nominal voltage is decreased; if the scale
parameter is increased, nominal voltage is also increased.
Range:
0.01 to
10.00
Default:
1.00
Increment:
0.01
Unit:
[
-
]
Scaling instrument B CMGZ421
Purpose:
With a setting of 1.00, the analogue output (0...5V) provides the nominal voltage at nominal
tension force. By decreasing the scale parameter, the nominal voltage is decreased; if the scale
parameter is increased, nominal voltage is also increased.
Range:
0.01 to
10.00
Default:
1.00
Increment:
0.01
Unit:
[
-
]