Operating Manual CMGZ411/421
23
11.1
Wiring diagram of the RS232 interface
Diagram: Wiring of the RS232 interface
C400007e
Reliable connection using maximum baudrate (9600) is guaranteed up to wire length of 10m. If the baudrate
is reduced and/or good conditions prevail, considerably greater distances can be bridged in some cases.
Connection to a PC etc. is done with a 9- or 25-pole Sub-D connector.
11.2
Command list
Command
Answer from controller
Purpose
IDNT<CR>
'CMGZ 411 V 1.00 0895'
10 characters type, fixed
< Typ > <Version> . . <S >
10 characters version, fixed
4 characters serial number, fixed
DAKT<CR>
XXXXX<CR>
read actual diameter (CMGZ411)
DIFF<CR>
DIFY<CR> / DIFN<CR>
alarm controller error
DIFR<CR>
DIFRXXXX.X<CR>
alarm controller error in %
ERR?<CR>
XXXXXX<CR>
ask for eventual errors
1
st
digit: 1: Err1, 0: no Err1
2
nd
digit: 1: Err2, 0: no Err2
3
rd
" " 3 " " 3
etc.
FREI<CR>
PACC<CR> / FAIL<CR>
enable controller
INRS<CR>
PACC<CR> / FAIL<CR>
interface initialization
(for ex. after loading new parameters)