Operating Manual CMGZ411/421
30
Value offset-feedback
Purpose:
This parameter allows to show the offset value in digits stored under „Find offset value“. It is not
needed to note this value, due to offset adjustment can be done easily at any time by using
parameter „Find offset value“.
Range:
-4000 to
+4000
Default:
0
Increment:
1
Value gain-feedback
Purpose:
This parameter allows to show the gain factor set under „Calibration feedback“ or to enter a gain
factor calculated by the formula (refer to „initial operation“).
Range:
0.100 to
9.000
Default:
1.000
Increment:
0.001
Value gain 1 feedback
Purpose:
This parameter allows to show the gain factor set under „Calibration gain 1 feedback“ or to enter a
gain factor calculated by the formula (refer to „initial operation“).
Range:
0.100 to
9.000
Default:
1.000
Increment:
0.001
Nominal force of sensors
Purpose:
To get a display value according to the actual force in the machine, it is necessary to enter the
nominal force of the sensors here.
Range:
1
to
9999
Default:
1000
Increment:
1
Unit:
[N, kN]
Unit force
Purpose:
This parameter stores the sensor’s force unit.
Range:
N
to
KN
Default:
N
Sensitivity
Purpose:
To get accurate force values in the display, the controller has to know the sensitivity of the force
sensor. FMS has 1.8mV/V as standard.
Range:
0.1
to
3.0
Default:
1.8
Increment:
0.1
Unit:
[mV/V]
Содержание CMGZ411
Страница 5: ...Operating Manual CMGZ411 421 5 3 3 Block Diagram CMGZ411 C411000e 3 4 Block Diagram CMGZ421 C421000e ...
Страница 9: ...Operating Manual CMGZ411 421 9 7 2 Determination of the machine configuration C400001e ...
Страница 10: ...Operating Manual CMGZ411 421 10 C400002e ...
Страница 11: ...Operating Manual CMGZ411 421 11 C400003e ...
Страница 12: ...Operating Manual CMGZ411 421 12 C400004e ...
Страница 13: ...Operating Manual CMGZ411 421 13 C400005e ...