Problem
Possible Cause
Corrective Action
Tidal volume errors in
expiratory only mode.
Some ventilators "servo"
control the base flow during
the expiratory period. Since
the
VT PLUS HF
must
assume the base flow is
constant during expiration,
errors will result due to the
changing base flow from
the ventilator during the
expiratory phase.
Use the bi-directional breath
detect mode. Ensure that the
VT PLUS HF
is placed
between the Y of the
breathing circuit and a test
lung.
Flow or pressure
measurement is not equal
to zero when zero flow or
pressure is applied.
Baseline drift of pressure
transducer
Zero the affected signal by
using the "zero" key while
the affected signal is being
displayed. For example, low
pressure can be zeroed
while the low-pressure
screen is currently displayed.
No message appears to
remove flow when
attempting to zero flow.
Appears that system will
not zero flow. No
message appears.
Autozeroing mode is
enabled.
VT PLUS HF
is zeroing flow
but no message is
supposed to be displayed
since the
VT PLUS HF
uses valves to switch to the
unused flow port.
This is normal. Since the
autozero mode is enabled,
the
VT PLUS HF
assumes
that the unused flow sensor
can be used to obtain the
zero reading.
Ensure the unused flow
sensor has zero flow when
the autozero mode is
enabled.
Zeroing flow results in
unexpected flow readings
or an error message.
Flow was not zero.
Ensure that there is no flow
when zeroing the flow
sensor.
Autozero mode was
enabled and the "unused"
flow sensor had a non-zero
flow present.
Whenever the autozero
mode is enabled, ensure that
there is no flow on the
unused flow sensor.
Barometric pressure is
incorrect
Barometric pressure offset
drift
Enter correct barometric
pressure.
No flow measurement is
observed when flow is
applied.
Wrong flow range is
selected
From the flow screen, ensure
that the range (low vs. high
flow) matches the range
depicted on the side panel of
the
VT PLUS HF
for the port
where the flow is being
applied.
Содержание VT Plus HF
Страница 2: ......
Страница 22: ...VT Plus HF Operators Manual 1 10 ...
Страница 64: ...VT Plus HF Operators Manual 5 16 ...
Страница 68: ...VT Plus HF Operators Manual 6 4 TREND TEST REPORT ...
Страница 71: ...Gas Flow Analyzer 6 6 7 LEAK TEST REPORT Fluke Biomedical Corporation ...
Страница 103: ...Gas Flow Analyzer VT Plus PC Software 8 8 13 Figure 8 7A Trend Test screen Figure 8 7B Trend Test screen ...
Страница 104: ...VT Plus HF Operators Manual 8 14 Figure 8 7C Trend Test screen ...
Страница 106: ...VT Plus HF Operators Manual 8 16 Figure 8 8A Leak Test screen Figure 8 8B Leak Test screen ...
Страница 108: ...VT Plus HF Operators Manual 8 18 Figure 8 9A Stacked Volume Test screen Figure 8 9B Stacked Volume Test screen ...
Страница 126: ...VT Plus HF Operators Manual 8 36 ...
Страница 134: ...VT Plus HF Operators Manual A 8 ...