6-1
Chapter 6
Special Functions
1.
Trend Test
The trend test is used to examine a single breath parameter over an extended period of
time. The trend test checks the parameter to see if it deviates beyond the limits set by the
user. The limits are set as a percent variation from the initial measured value. There are
two ways to access the trend screen:
1.
Press the
FULL/5
key to display the breath parameters.
2.
Press the soft key below
TREND.
Or
1.
Press the
MORE/4
key.
2.
Use the
∧
and
∨
soft keys to highlight
TREND TEST
.
3.
Press the soft key below
ENTER.
After selecting
TREND TEST
, the
TREND TEST
setup screen is displayed.
Trend Test
Leak Test
Stacked Volume Test
RT200 Emulation Mode
Assist Test
High-Frequency Oscillator Mode
Содержание VT Plus HF
Страница 2: ......
Страница 22: ...VT Plus HF Operators Manual 1 10 ...
Страница 64: ...VT Plus HF Operators Manual 5 16 ...
Страница 68: ...VT Plus HF Operators Manual 6 4 TREND TEST REPORT ...
Страница 71: ...Gas Flow Analyzer 6 6 7 LEAK TEST REPORT Fluke Biomedical Corporation ...
Страница 103: ...Gas Flow Analyzer VT Plus PC Software 8 8 13 Figure 8 7A Trend Test screen Figure 8 7B Trend Test screen ...
Страница 104: ...VT Plus HF Operators Manual 8 14 Figure 8 7C Trend Test screen ...
Страница 106: ...VT Plus HF Operators Manual 8 16 Figure 8 8A Leak Test screen Figure 8 8B Leak Test screen ...
Страница 108: ...VT Plus HF Operators Manual 8 18 Figure 8 9A Stacked Volume Test screen Figure 8 9B Stacked Volume Test screen ...
Страница 126: ...VT Plus HF Operators Manual 8 36 ...
Страница 134: ...VT Plus HF Operators Manual A 8 ...