Gas Flow Analyzer
5
5-13
7.
Full Breath Parameters Test Screen
The full test screen is used to display all of the breath parameters calculated by the
VT
PLUS HF
. Note that breath detection must be enabled for breath parameters to be
calculated. See
Chapter 4, System Setup
for instructions on enabling breath detection.
Trend:
Press the soft key below
TREND
to obtain a printed trend of a single
parameter. See Chapter 6, Special Functions for a more detailed
description of the trend function.
Clear:
Press the soft key below
CLEAR
to reset the statistics and re-start the
plot.
When an asterisk appears for a parameter value, the parameter is either invalid for the
current
breath detect mode
or the system does not have sufficient information to
accurately calculate the parameter. For example, base flow is invalid during bi-directional
breath detection mode since the base flow is only seen in the inspiratory and expiratory
limbs of the ventilator circuit. In another example, it would be impossible to measure the
inspiratory hold pressure during the expiratory-only mode.
Note:
The Assist Pressure "Assist" and the Minimum Pressure "Min. Pres" are
not available for trending and cannot be added to other screens using the
parameter selection. Refer to Chapter 6 for instructions on Assist Mode
testing.
Содержание VT Plus HF
Страница 2: ......
Страница 22: ...VT Plus HF Operators Manual 1 10 ...
Страница 64: ...VT Plus HF Operators Manual 5 16 ...
Страница 68: ...VT Plus HF Operators Manual 6 4 TREND TEST REPORT ...
Страница 71: ...Gas Flow Analyzer 6 6 7 LEAK TEST REPORT Fluke Biomedical Corporation ...
Страница 103: ...Gas Flow Analyzer VT Plus PC Software 8 8 13 Figure 8 7A Trend Test screen Figure 8 7B Trend Test screen ...
Страница 104: ...VT Plus HF Operators Manual 8 14 Figure 8 7C Trend Test screen ...
Страница 106: ...VT Plus HF Operators Manual 8 16 Figure 8 8A Leak Test screen Figure 8 8B Leak Test screen ...
Страница 108: ...VT Plus HF Operators Manual 8 18 Figure 8 9A Stacked Volume Test screen Figure 8 9B Stacked Volume Test screen ...
Страница 126: ...VT Plus HF Operators Manual 8 36 ...
Страница 134: ...VT Plus HF Operators Manual A 8 ...