VT Plus HF
Operators Manual
A-4
2. General
Specifications
Operating Environmental Conditions
Temperature Range............................................................................................ 10 to 40
°
C
Ambient Humidity 0 - 80% non-condensing to 31
°
C, decreasing to 50% at 40
°
C
Barometric
Pressure ...........................................................................................8 to 18 psia
Vibration......................... Device is intended to be used on tabletop or other stable surface.
Storage Environmental Conditions
Temperature Range........................................................................................... -25 to 50
°
C
Humidity...................................................................................0 to 95% non-condensing
°
C
Power Requirements
Maximum Over-voltage...........................................................................................264 VAC
Input
Voltage
Range .................................................................................... 100 to 240 VAC
Input
Frequency
Range .......................................................................................... 50/60 Hz
Power
Consumption...............................................................................< 132 Volt-Amperes
Fuse
Rating .................................................................................................... 0.5 A, Slo-Blo
3. Performance
Specifications
Low-Pressure Port
Maximum Applied Pressure .........................................................................................60 psi
Operating Pressure (Differential) ...................................................... . ±500 mmHg (
±
10 psi)
Operating
Pressure
(Common-mode)......................................................................... 30 psi
Span
Accuracy ............................
±
0.50% of reading or
±
1.5 mmHg, whichever is greater
Frequency
Response................................................................................................> 10 Hz
Resolution.............................................................................................................0.1 mmHg
Sample
Rate ..............................................................................................................100 Hz
Note:
Fluid pressure may be applied to the positive port; however, fluids should be kept from entering the
pressure port by using a suitable length of connection tubing.
High-Pressure Port
Maximum
Applied
Pressure .......................................................................................150 psi
Operating
Pressure.................................................................................................
±
100 psi
Span
Accuracy .....................................
±
1% of reading or
±
0.1 psig, whichever is greater
Frequency
Response................................................................................................> 10 Hz
Resolution...................................................................................................................0.1 psi
Sample
Rate ..............................................................................................................100 Hz
Содержание VT Plus HF
Страница 2: ......
Страница 22: ...VT Plus HF Operators Manual 1 10 ...
Страница 64: ...VT Plus HF Operators Manual 5 16 ...
Страница 68: ...VT Plus HF Operators Manual 6 4 TREND TEST REPORT ...
Страница 71: ...Gas Flow Analyzer 6 6 7 LEAK TEST REPORT Fluke Biomedical Corporation ...
Страница 103: ...Gas Flow Analyzer VT Plus PC Software 8 8 13 Figure 8 7A Trend Test screen Figure 8 7B Trend Test screen ...
Страница 104: ...VT Plus HF Operators Manual 8 14 Figure 8 7C Trend Test screen ...
Страница 106: ...VT Plus HF Operators Manual 8 16 Figure 8 8A Leak Test screen Figure 8 8B Leak Test screen ...
Страница 108: ...VT Plus HF Operators Manual 8 18 Figure 8 9A Stacked Volume Test screen Figure 8 9B Stacked Volume Test screen ...
Страница 126: ...VT Plus HF Operators Manual 8 36 ...
Страница 134: ...VT Plus HF Operators Manual A 8 ...