![Fluke VT Plus HF Скачать руководство пользователя страница 80](http://html1.mh-extra.com/html/fluke/vt-plus-hf/vt-plus-hf_operators-manual_2310039080.webp)
VT Plus HF
Operators Manual
7-2
2.
Replacing the Oxygen Sensor
The
VT PLUS HF
uses a galvanic oxygen sensor that is warranted for 12 months. The
oxygen sensor cell needs to be replaced when the oxygen calibration cannot be
successfully completed. The oxygen sensor may operate for longer than 12 months
depending on use. To replace the oxygen sensor, follow these steps below.
Disconnect the power cord before opening the oxygen sensor door.
1.
Remove the three screws in the oxygen sensor door on the back of the
VT PLUS HF
.
The oxygen sensor is a round plastic cylinder about one inch in diameter with wires
protruding from the back of the sensor.
2.
Disconnect the in-line electrical connector on the sensor wire.
3.
Un-screw the old sensor and remove it.
4.
Screw in the new sensor (Fluke Biomedical P/N 2138514).
5.
Re-connect the sensor wire.
6.
Replace the oxygen sensor door.
7.
Re-calibrate the oxygen sensor following the instructions below.
3.
Procedure for Calibrating the Oxygen
Enter the oxygen calibration screen as follows:
1.
Enter the setup menu by pressing the
SETUP
key.
2.
Press the ^ and v soft keys to select
UTILITIES
and then press the
ENTER
soft key.
3.
Press the ^ and v soft keys to select
OXYGEN CALIBRATION
and then press the
ENTER
soft key.
Note:
When
AIRWAY
is selected, the airway transducer will be connected to the
currently selected.
Содержание VT Plus HF
Страница 2: ......
Страница 22: ...VT Plus HF Operators Manual 1 10 ...
Страница 64: ...VT Plus HF Operators Manual 5 16 ...
Страница 68: ...VT Plus HF Operators Manual 6 4 TREND TEST REPORT ...
Страница 71: ...Gas Flow Analyzer 6 6 7 LEAK TEST REPORT Fluke Biomedical Corporation ...
Страница 103: ...Gas Flow Analyzer VT Plus PC Software 8 8 13 Figure 8 7A Trend Test screen Figure 8 7B Trend Test screen ...
Страница 104: ...VT Plus HF Operators Manual 8 14 Figure 8 7C Trend Test screen ...
Страница 106: ...VT Plus HF Operators Manual 8 16 Figure 8 8A Leak Test screen Figure 8 8B Leak Test screen ...
Страница 108: ...VT Plus HF Operators Manual 8 18 Figure 8 9A Stacked Volume Test screen Figure 8 9B Stacked Volume Test screen ...
Страница 126: ...VT Plus HF Operators Manual 8 36 ...
Страница 134: ...VT Plus HF Operators Manual A 8 ...