VT Plus HF
Operators Manual
1-4
Oxygen Measurements
The oxygen measurement is performed using a galvanic oxygen cell on the high-flow
port near the exhaust.
The
VT PLUS HF
interprets the measurement from the oxygen
sensor as a partial pressure of the total pressure in the high-flow circuit.
The total
pressure is the airway pressure plus the barometric pressure.
The oxygen percentage is calculated as the fraction of the partial pressure of oxygen
relative to the total pressure.
It should be noted that changes in the total pressure (i.e.
airway pressure) will result in a change in the partial pressure of the measured oxygen
even though the percentage of oxygen is the same.
Since the pressure can change much
faster than the response time of the oxygen sensor, a small transient (1 to 2% of full-
scale) error can occur in the oxygen measurement when there is a fast change in the
pressure at the oxygen sensor.
Whenever an oxygen measurement is made, or when the oxygen sensor is calibrated or
zeroed, there should be sufficient gas flow pass the oxygen sensor.
Only a small flow of
gas (say, 10 LPM)
is needed to ensure that the oxygen concentration at the oxygen sensor
is as expected.
For example, just opening the high-flow exhaust to room air does not
ensure that the oxygen sensor will "see" 20.9% oxygen.
The oxygen concentration at the
sensor will not equilibrate with the room air unless a small amount of flow is present to
ensure gas mixing.
The oxygen sensor should last for more than one year, and is replaced during an annual
calibration of the unit. Sensor life is shortened, the longer the sensor is exposed to
oxygen. This includes oxygen in room air as well as higher flow rates of oxygen-
containing gases past the sensor.
The
VT PLUS HF
will check to detect if the oxygen
sensor is depleted every time an oxygen zero is performed.
The oxygen sensor should be recalibrated by the user with 100% oxygen and dry air at
the beginning of any testing day in which oxygen concentrations will be measured.
Refer
to
Chapter 7, Safety, Error Messages, Troubleshooting, and Support
for more details.
Flow Measurements
The flow measurements in the
VT PLUS HF
are made by measuring the differential
pressure across a resistive screen mesh.
The resulting differential pressure is nearly linear
with respect to the volumetric flow.
Simple corrections are applied to correct for any non-
linearities.
A table of correction factors is applied for each the low- and high-flow
sensors.
The information regarding these tables can be seen using the
SETUP | UTILITES
| LINEARIZATION
screen on the
VT PLUS HF
.
Gas Settings
: The flow measurement sensors in the
VT PLUS HF
are linearly dependent
on the viscosity of the applied gas.
The
VT PLUS HF
setup options allow the user to
specify what type of gas is being used.
Additional choices are available using the USER
DEFINE MODE.
The USER DEFINED gas settings are set using the PC software.
The
gas calculator on the PC software can determine the effective viscosity for a combination
of various gases.
Refer to
Chapter 8, VT Plus PC Software
for more details on the gas
calculator and on setting user defined gas types.
The
VT PLUS HF
calculates the temperature dependent viscosity for the gas type selected
(except for user-defined gases).
Therefore, when a new flowing gas temperature is
Содержание VT Plus HF
Страница 2: ......
Страница 22: ...VT Plus HF Operators Manual 1 10 ...
Страница 64: ...VT Plus HF Operators Manual 5 16 ...
Страница 68: ...VT Plus HF Operators Manual 6 4 TREND TEST REPORT ...
Страница 71: ...Gas Flow Analyzer 6 6 7 LEAK TEST REPORT Fluke Biomedical Corporation ...
Страница 103: ...Gas Flow Analyzer VT Plus PC Software 8 8 13 Figure 8 7A Trend Test screen Figure 8 7B Trend Test screen ...
Страница 104: ...VT Plus HF Operators Manual 8 14 Figure 8 7C Trend Test screen ...
Страница 106: ...VT Plus HF Operators Manual 8 16 Figure 8 8A Leak Test screen Figure 8 8B Leak Test screen ...
Страница 108: ...VT Plus HF Operators Manual 8 18 Figure 8 9A Stacked Volume Test screen Figure 8 9B Stacked Volume Test screen ...
Страница 126: ...VT Plus HF Operators Manual 8 36 ...
Страница 134: ...VT Plus HF Operators Manual A 8 ...