Gas Flow Analyzer
5
5-9
Low Pressure Screen
The low-pressure screen is used to display the signal and statistics for the pressure signal
from the low-pressure ports of the
VT PLUS HF
.
Range:
Press the soft key below
RANGE
to change between low pressure,
airway pressure, and high-pressure screens.
Units:
Press the soft key below
UNITS
to change the units of measure for the
low-pressure screen. Possible units of measure for this screen include:
mbar:
millibar = .001 BAR
bar:
14.7 PSIG (Barometric)
mmHg:
millimeters of mercury.
inHg:
inches
of
mercury.
cmH
2
O:
centimeters of water.
inH
2
O:
inches of water.
PSIG:
pounds per square inch
atm:
atmospheres
kPa:
kilopascals
Re-scale:
Press the soft key below
RESCALE
to change the scale on the plot of
data to a more appropriate window of values.
Note:
RESCALE
uses the current Min/Max values to determine plot limits.
Clear:
Press the soft key below
CLEAR
to reset the statistics and re-start the
plot.
The user can select the breath parameters displayed in this screen. Follow the instructions
on page 5-5 to select which breath parameters are displayed in this screen.
Содержание VT Plus HF
Страница 2: ......
Страница 22: ...VT Plus HF Operators Manual 1 10 ...
Страница 64: ...VT Plus HF Operators Manual 5 16 ...
Страница 68: ...VT Plus HF Operators Manual 6 4 TREND TEST REPORT ...
Страница 71: ...Gas Flow Analyzer 6 6 7 LEAK TEST REPORT Fluke Biomedical Corporation ...
Страница 103: ...Gas Flow Analyzer VT Plus PC Software 8 8 13 Figure 8 7A Trend Test screen Figure 8 7B Trend Test screen ...
Страница 104: ...VT Plus HF Operators Manual 8 14 Figure 8 7C Trend Test screen ...
Страница 106: ...VT Plus HF Operators Manual 8 16 Figure 8 8A Leak Test screen Figure 8 8B Leak Test screen ...
Страница 108: ...VT Plus HF Operators Manual 8 18 Figure 8 9A Stacked Volume Test screen Figure 8 9B Stacked Volume Test screen ...
Страница 126: ...VT Plus HF Operators Manual 8 36 ...
Страница 134: ...VT Plus HF Operators Manual A 8 ...