Step 1: Set the Reference in Smart Remote Mode
1-1
Turn on the tester and remote and let them sit for a
minimum of 5 minutes. Let them sit longer if they are above
or below ambient temperature.
1-2
Make sure that the home screen shows the correct settings for
the job, and the test type is
Smart Remote
.
To make sure that other settings are correct, tap the test
setup panel, make sure the correct test is selected on the
CHANGE TEST
screen, then tap
EDIT
to see more settings.
Table 6 on page 111 describes the settings.
1-3
Clean and inspect the connectors on the tester, remote, and
test reference cords.
1-4
On the home screen tap
SET REF
.
1-5
On the
SET REFERENCE
screen, tap
RUN WIZARD
.
Notes
To only set the reference, and not measure the
loss of your test reference cords, tap SKIP WIZARD
on the SET REFERENCE screen.
Fluke Networks recommends that you measure
the loss of your test reference cords each time you
set the reference.
1-6
Make the connections to set the reference, as shown on the
screen and in Figure 47, then tap
NEXT
to see the completed
connections.
Notes
The SET REFERENCE screen shows reference
connections for the selected reference method.
Figure 47 shows connections for a 1 Jumper
reference.
When you set the reference, align the testers as
shown in Figure 47 to keep the fibers as straight as
possible.
-continued-
Содержание Versiv
Страница 38: ...Versiv Cabling Certification Product Family Users Manual 24...
Страница 86: ...Versiv Cabling Certification Product Family Users Manual 72...
Страница 202: ...Versiv Cabling Certification Product Family Users Manual 188...
Страница 214: ...Versiv Cabling Certification Product Family Users Manual 200...
Страница 234: ...Versiv Cabling Certification Product Family Users Manual 220...
Страница 244: ...Versiv Cabling Certification Product Family Users Manual 230...
Страница 256: ...Versiv Cabling Certification Product Family Users Manual 242...
Страница 258: ...Versiv Cabling Certification Product Family Users Manual 244...