4. REMOTE OPERATION
Page 123
© 1998-2007 DH Instruments, a Fluke Company
STDRES(=)
Purpose
Read or set the molbox’s two internal standard resistor values.
Syntax “STDRES”
“STDRES=R0
100
, R0
110
Arguments R0
100
:
R0
110
:
The 100 ohm standard resistor value (ohms).
The 110 ohm standard resistor value (ohms).
Remarks
The standard resistor values are used to periodically calibrate the
PRT measurement system. These values are stored in the
molbox’s factory data area, and should be modified with care.
You should not routinely modify these values, as you risk data
corruption in doing so. After modifying these values, a calibration
of the PRT measurement system will occur and an updated
measurement of the PRTs will not be available until this is
complete (about 20 seconds)
Example Command:
Reply:
“STDRES=100.0022,110.0132”
“100.0022 Ohms, 110.0132 Ohms”
Errors
ERR# 6:
Values are invalid (must be between 1 and 199).
See Also
“OHMS”, “TEMP”
TARE
Purpose
Read the current
tare
conditions and the current
tare
.
Syntax “TARE”
Remarks
The user should set the molbox RFM valves to the
tare
condition
(see “TARESETUP”) and check the
tare
conditions before taring
the internal transducers (see “TARESET”). The returned data will
allow the user to determine if the current conditions will allow a
valid
tare
. The reply contains three or five fields which are
separated by commas.
The first field starts with an "R" if the system is ready to
tare
or
else an "NR". The "R" must be present to
tare
the transducers.
The system is ready to
tare
only if the current difference in
pressure [Pa] between the up and down stream RPTs without
tare
is below 9999 Pa. If the microrange option is present, then
the microrange RPT pressure without
tare
must be below
999 Pa. The actual pressure rate of change in Pa/sec
follows
this text.
The second field displays the current difference in pressure [Pa]
between the up and down stream RPTs without
tare
. The third
field shows the last
tare
value.
If the optional microrange option is present, then the microrange
RPT pressure [Pa] without
tare
will appear in a fourth field, and
the fifth field shows the last microrange
tare
value. All returned
values are in Pascal [Pa].
Example Command:
Reply:
“TARE”
“R 0 Pa/s, 115 Pa, 108 Pa” (no microrange)
Command:
Reply:
“TARE”
“R 0 Pa/s, 115 Pa, 108 Pa, 6 Pa, 3 Pa (with
microrange)
See Also
3.2.4.1, “TARESET”, “TARESETUP”
Содержание molbox RFM
Страница 10: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S ...
Страница 142: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 132 N NO OT TE ES S ...
Страница 158: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 148 N NO OT TE ES S ...
Страница 164: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 154 N NO OT TE ES S ...
Страница 170: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 160 N NO OT TE ES S ...