4. REMOTE OPERATION
Page 105
© 1998-2007 DH Instruments, a Fluke Company
ABORT
Purpose
To stop an active averaging,
tare
, purge or leak check cycle.
Syntax “ABORT”
Remarks
This command should be used to ensure that molbox RFM is in
a known state. The
tare
, purge, and leak check cycles put the
RFM into a condition where it is not usable for normal flow
measurement, so the “ABORT” command should be used
before general operation of the unit begins if the previous state
of the unit is not known. The “ABORT” command also clears out
any previous leak or
tare
cycle errors.
Example Command:
Reply:
“ABORT”
“ABORT”
See Also
“FA”, “TARESETUP”, “LEAKCK”, “PURGE”
AIRW(=)
Purpose
Read or set the humidity ratio for the gas, AIR.
Syntax “AIRW=
HumidityRatio
”
“AIRW”
Default “AIRW=0.00
00”
Arguments
HumidityRati
o:
The humidity ratio (0-0.1)
Remarks
The humidity ratio is only used if the gas is set to specify AIR. For all
other gas types, the ratio is ignored.
Example Command:
Reply:
“AIRW=.01”
“0.0100”
Errors
ERR# 6:
The humidity ratio is invalid.
See Also
3.2.2, 3.4.2.2
“GAS”
Содержание molbox RFM
Страница 10: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S ...
Страница 142: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 132 N NO OT TE ES S ...
Страница 158: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 148 N NO OT TE ES S ...
Страница 164: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 154 N NO OT TE ES S ...
Страница 170: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 160 N NO OT TE ES S ...