molbox™ RFM™ OPERATION AND MAINTENANCE MANUAL
© 1998-2007 DH Instruments, a Fluke Company
Page 10
Table 4.
molbloc-S Flow at Various molbloc Upstream Pressures
FLOW [slm @ 0°C] WHEN molbloc-S UPSTREAM PRESSURE IS:
[1][2]
molbloc-S
DESIGNATO
R
K
F
[sccm/kPa]
20 kPa
(3psia)
50 kPa
(7 psia)
100 kPa
(15 psia)
Minimum
without
vacuum
[3
]
150 kPa
(22 psia)
200 kPa
(30 psia)
250 kPa
(36 psia)
500 kPa
(70 psia)
5E1-S
50 1 2.5 5 7.7 7.5 10
12.5 25
1E2-S
100
2
5
10
15
15
20
25
50
2E2-S
200 4 10 20 28 30 40 50
100
5E2-S
500
10
25
50
67
75
100
125
250
1E3-S
1
000 20 50 100 129 150 200 250 500
2E3-S
2 000
40
100
200
248
300
400
500
1 000
5E3-S
5
000 100 250 500 596 750
1
000
1
250
2500
1E4-S
10 000
200
500
1 000
1 173
1 500
2 000
2 500
5 000
[1] When volumetrically based mass flow units with reference temperatures other than 0°C are used, flow values
will generally be higher; for example, the flow values for a given molbloc and upstream pressure are
approximately 7% higher when expressed in slm @ 20°C. Flow values at a given pressure may vary by up to
2% due to flowpath machining tolerances.
[2] Flow values in table are valid only when critical flow is established.
[3] Minimum upstream pressure to achieve critical flow with atmospheric pressure (approximately 100 kPa)
downstream of molbloc-S (no vacuum).
Table 5.
Minimum molbloc-S Critical Flow (slm) at Various molbloc-S Downstream Pressures
MINIMUM molbloc-S CRITICAL FLOW [SLM @ 0°C] WITH molbloc DOWNSTREAM PRESSURE OF:
[1]
molbloc-S
DESIGNATO
R
≤
5
kPa
(0.7
psia)
10
kPa
(1.5
psia)
25
kPa
(3.5
psia)
50
kPa
(7
psia)
100
kPa
(15
psia)
110
kPa
(16
psia)
125
kPa
(18
psia)
150
kPa
(22
psia)
200
kPa
(30
psia)
250
kPa
(36
psia)
300
kPa
(44
psia)
5E1-S
1
[2]
1.7 3.4 4.7 7.7 8.4 9.4 11 14 17 20
1E2-S
2
[2]
3.4
5.9
8.4
15
16
18
21
27
33
38
2E2-S
4
[2]
5.9 9.8 16 28 31 34 40 51 63 74
5E2-S
10
[2]
12
20
37
67
72
80
95
122
149
179
1E3-S
20
[2]
20
[2]
39 69 129 139 154 184 239 294 349
2E3-S
40
[2]
40
[2]
73
131
248
268
298
358
468
578
687
5E3-S
100
[2]
100
[2]
173 317 596 646 746 871
1
145
1
420
1
694
1E4-S
200
[2]
200
[2]
347
615
1 173
1 273
1 442
1 741
2 240
2 789
3 338
[1] When volumetrically based mass flow units with reference temperatures other than 0 °C are used, flow values
will generally be higher; for example, the flow values for a given molbloc and upstream pressure are
approximately 7 % higher when expressed in slm @ 20 °C. Flow values at a given pressure may vary by up to
2 % due to flowpath machining tolerances.
[2] Limited by 20 kPa minimum calibration pressure rather than back pressure ratio.
Содержание molbox RFM
Страница 10: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S ...
Страница 142: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 132 N NO OT TE ES S ...
Страница 158: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 148 N NO OT TE ES S ...
Страница 164: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 154 N NO OT TE ES S ...
Страница 170: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 160 N NO OT TE ES S ...