4. REMOTE OPERATION
Page 117
© 1998-2007 DH Instruments, a Fluke Company
PCALDATE(=)
Purpose
Read or set the user RPT calibration dates.
Syntax
“PCALDATE=yyyymmdd, yyyymmdd (,yyyymmdd)”
“PCALDATE”
Defaults
“PCALDATE=19980101,19980101 (no microrange option)
“PCALDATE=19980101,1998010119980101” (microrange
option)
Arguments
yyyy:
mm:
dd:
The year from 1980 to 2079
The month from 1 to 12
The day from 1 to the last valid day of the given
month
Remarks
The user defined pressure calibration dates for the upstream and
downstream RPTs (and optional microrange RPT) can be access
with this single command in this order.
Example Command:
Reply:
“PCALDATE=19981005, 19981005” (no
microrange option)
“19981005, 19981005”
Command:
Reply:
“PCALDATE=19981005,19981005,19981005”
(microrange option)
“19981005,19981005,19981005”
Errors
ERR# 7:
One or more of the date(s) are invalid.
See Also
5.2, “PCAL(=)”
PCOEF
Purpose
Read the coefficient to convert Pascal [Pa] to the current
pressure units.
Syntax “PR”
Remarks
The pressure coefficient (PCOEF) is a value that is used to convert
Pascal units to the current pressure units. To use this coefficient,
multiply it by pressure in Pascal to get pressure in the current units.
Example Command:
Reply:
“PCOEF”
“1.000000e-003”
See Also
7.1.1, “UNIT”
PR
Purpose
Read the next molbloc average pressure.
Syntax “PR”
Remarks
The next available average
upstream downstream
−
⎛
⎝⎜
⎞
⎠⎟
2
pressure value is
read in the current pressure units. It is the average of the tare
corrected upstream and the downstream transducers. The data
string also contains flow ready information and the pressure units.
The first 3 characters of the reply are reserved for the flow
ReadyNot Ready status. The Ready/Not Ready status is covered in
the “SR” command. The pressure measurement number starts at
the fifth character, and is followed by the pressure unit of measure.
Example Command:
Reply:
“PR”
“R 247.598 kPaa”
See Also
3.1.2.4, “READYCK”, “SR”, “PRDP”, “PRHI”, “PRLO”
Содержание molbox RFM
Страница 10: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S ...
Страница 142: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 132 N NO OT TE ES S ...
Страница 158: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 148 N NO OT TE ES S ...
Страница 164: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 154 N NO OT TE ES S ...
Страница 170: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 160 N NO OT TE ES S ...