molbox™ RFM™ OPERATION AND MAINTENANCE MANUAL
© 1998-2007 DH Instruments, a Fluke Company
Page 38
To measure flow in volume flow units (sometimes referred to as
actual
flow
units), press
[UNIT]
,
<vlm>
under the UNIT
function. If
<vlm>
is not available,
see Section 3.4.3.5. This selection accesses the menu of volume units available.
Select the desired volume flow unit:
1.
Current volume flow pressure.
2. Current volume flow temperature. Indicates
<bloc>
if the current setting is to use the
molbloc temperature.
P101.325kPa T21.1ºC
Edit P&T? 1no 2yes
To retain the current volume flow pressure and temperature, select
<1no>
and
operation returns to the MAIN run screen with the selected volume flow unit
active. To change the volume flow pressure and/or temperature select
<2yes>
.
The next screen gives the choice of
having the volume flow gas temperature
be either the molbloc temperature or a
user specified temperature. If
<2user>
is selected, a screen to enter the temperature
is presented before continuing. If
<1molbloc>
is selected, the volume flow
temperature will automatically be taken
as the molbloc temperature.
Gas temperature:
1molbloc 2user
The next screen is to edit the volume
flow pressure in the current pressure unit
of measure. Pressing
[ENTER]
accepts
the edited value as the volume flow
pressure and returns to the MAIN run
screen with the selected volume flow unit
and gas temperature and pressure active.
Volume unit gas pres
101.325 kPa
The temperature and pressure units of measure used to specify volume flow
conditions can be changed using [SETUP], <2presU> for pressure (see
Section 3.5.2) and [SETUP], <3tempU> for temperature (see Section 3.5.3).
Because volume flow is dependent on gas pressure and temperature at the
flow measurement point, gas pressure and temperature must be specified by
the user when selecting volume flow units. The measurement uncertainty in
the volume flow measurement is highly dependent on the measurement
uncertainty in the pressure and temperature specified (see Section 3.4.3.1).
3.4.3.5
CUSTOMIZING FLOW UNITS AVAILABLE UNDER THE
UNIT FUNCTION
The UNIT function provides a choice of six different flow units of measure. The units
that are available by default are the six indicated in Section 3.4.3. However,
molbox RFM supports many other units. These other units can be made
available for selection by customizing the UNIT function.
To customize the UNIT function,
press
[SETUP]
and select
<1flowU>
.
The display is:
Set up user unit #1
Содержание molbox RFM
Страница 10: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S ...
Страница 142: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 132 N NO OT TE ES S ...
Страница 158: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 148 N NO OT TE ES S ...
Страница 164: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 154 N NO OT TE ES S ...
Страница 170: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 160 N NO OT TE ES S ...