molbox™ RFM™ OPERATION AND MAINTENANCE MANUAL
© 1998-2007 DH Instruments, a Fluke Company
Page 100
Table 22.
Error Messages
REPLY
DESCRIPTION
ERROR
“The molbox RFM is in high security level and cannot accept remote commands”
ERR# 0
“OK”
ERR# 1
“molbloc flow is too great”
ERR# 2
“Text argument is too long”
ERR# 3
“User defined coefficient cannot be 0”
ERR# 4
“External device not detected”
ERR# 5
“External device improperly configured”
ERR# 6
“Numeric argument missing or out of range”
ERR# 7
“Missing or improper command argument(s)”
ERR# 8
“External device timeout error”
ERR# 9
“Unknown command”
ERR# 11
“Command missing argument”
ERR# 12
“System overpressured”
ERR# 13
“Text detected in numeric field”
ERR# 14
“User unit not defined”
ERR# 15
“Averaging cycle not started”
ERR# 17
“Selected gas not available”
ERR# 18
“Command not yet available”
ERR# 19
“Not supported”
ERR# 20
“Not supported”
ERR# 21
“User device not defined”
ERR# 22
“Pressure is not stable”
ERR# 23
“Option not available or installed”
ERR# 24
“molbloc not detected”
ERR# 25
“Transducer out of calibration”
ERR# 26
“COM port failed to initialize”
ERR# 27
“Internal device #1, 2 or 3 timeout error’
ERR# 29
“Busy averaging”
ERR# 30
“Fatal mass flow calculation error”
ERR# 31
“molbloc EEPROM is full”
ERR# 32
“molbloc gas not found”
ERR# 34
“molbloc is write protected”
ERR# 35
“molbloc write error”
ERR# 37
“Gas table is full”
ERR# 38
“Selected range not available”
ERR# 39
“molbloc verify error”
ERR# 40
“molbloc read error”
ERR# 41
“molbloc invalid”
ERR# 42
“PRT measurement error”
ERR# 44
“Entry already exists”
ERR# 45
“Argument not allowed”
ERR# 46
“The leak is too large”
ERR# 47
“Temperature change was too great”
ERR# 48
“Offset was excessive. Check
tare
”
ERR# 49
“Possible upstream leak”
ERR# 50
“Possible downstream leak”
Содержание molbox RFM
Страница 10: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S ...
Страница 142: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 132 N NO OT TE ES S ...
Страница 158: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 148 N NO OT TE ES S ...
Страница 164: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 154 N NO OT TE ES S ...
Страница 170: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 160 N NO OT TE ES S ...