
molbox™ RFM™ OPERATION AND MAINTENANCE MANUAL
© 1998-2007 DH Instruments, a Fluke Company
Page 110
FR
Purpose
Read the next measurement of flow through the molbloc as
calculated by the molbox RFM.
Syntax “FR”
Remarks
The next available flow value is read in the current flow units.
This can take up to 1 second, since the molbox RFM waits for
the next flow measurement before replying. The reply also
contains ready information.
The first 3 characters are reserved for the
ReadyNot Ready
status. The
Ready/Not Ready
status is covered in the “SR”
command. The flow measurement number starts at the fifth
character, and is followed by the flow units.
Example Command:
Reply:
“FR”
“R 0.00001 sccm”
Errors
ERR# 27:
One of the RPTs (transducer) is not functioning.
See Also
3.1.2.4, “SR”, “READYCK”
FRA
Purpose
Read the results of a completed flow averaging cycle.
Syntax “FRA”
Remarks
This command is used to check on the status of or collect the
results of an averaging cycle that was started earlier using the
"FA=" command. During the averaging cycle, this command will
reply "BUSY" until the averaging cycle is complete. Then the
reply will be the result of the averaging cycle. If the cycle is
aborted before it is complete (see "FA="), then the data will be
lost.
The data that is returned from the command consists of multiple
numeric fields delimited by commas. The flow numbers are in
the current flow unit of measure. The actual data fields returned
are as follows:
An "H" will appear as the first character of the reply string. An
"S" will appear as the second character of the string if the flow
stability stayed within the stability setting during the sequence.
From the fourth position on, the numeric data fields are:
Average flow over the averaging period.
Standard deviation of flow over the averaging period.
The minimum flow reached during the period.
The maximum flow reached during the period.
“NA” always.
“NA” always.
Example Command:
Reply:
Reply:
“FRA”
“HS 0.00002 sccm,0.00000,0.00002,0.00002,NA,
NA”
“BUSY” if still averaging.
Errors
ERR# 15:
The “FR” command was not sent previous to
this command.
See Also
3.2.6.2, “FR”, “SR”, “FA”
Содержание molbox RFM
Страница 10: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S ...
Страница 142: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 132 N NO OT TE ES S ...
Страница 158: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 148 N NO OT TE ES S ...
Страница 164: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 154 N NO OT TE ES S ...
Страница 170: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 160 N NO OT TE ES S ...