3. OPERATION
Page 23
© 1998-2007 DH Instruments, a Fluke Company
3.1.4
SOFT [ON/OFF] KEY
molbox RFM is equipped with a SOFT
[ON/OFF]
key and indicator LED on the bottom left hand
corner of the front panel. The purpose of the SOFT ON/OFF key is to put molbox RFM into a
dormant mode in which the display is turned OFF but power is still supplied and OVERPRESSURE
functions are still active. When molbox RFM is ON, the ON/OFF indicator is ON continuously.
When molbox RFM is SOFT OFF, the ON/OFF indicator blinks every 5 seconds.
When molbox RFM is SOFT OFF, receiving a remote command turns it ON.
When molbox RFM is SOFT OFF, an overpressure conditions turns it ON.
3.1.5
MICRORANGE OPTION (OPTIONAL)
The molbox RFM offers a microrange option to improve molbloc-L flow measurement specifications
below 10 % of full scale of the molbloc being used (see Section 1.2.4.1.1). The microrange option is
only active during molbloc-L operation and has no effect on molbloc-S operation or specifications.
The microrange option includes a low differential pressure RPT to improve the measurement of
differential pressure across the molbloc below 12.5 kPa (1.8 psi). A three-way valve and on-board
logic automatically put the low differential RPT into and out of service. This allows it to be used as
the source of differential pressure values whenever differential pressure is under 12.5 kPa (1.8 psi)
and bypassed for protection from overpressure when differential pressure exceeds 13.5 kPa (2 psi)
(see Figure 3).
With automatic microrange operation ON, the microrange option (if present) is used transparently to
the operator to optimize flow measurement resolution and accuracy (see Section 3.4.7). With
automatic microrange operation OFF, the microrange option is not used and all measurements are
made using the upstream and downstream absolute RPTs only.
When the type of molbloc connected to molbox RFM is changed from molbloc-L to molbloc-S, or
molbox RFM power is cycled, the last state of the microrange option (ON or OFF) is retained for
molbloc-L operation.
The microrange option can also be controlled manually by pressing
[SPECIAL]
and selecting
<7micro>
(see Section 3.6.7).
The microrange indicator on the main run screen (see Section 3.4.7) indicates the status of the
microrange option.
1.
High Isolation:
Open
2. Low
Isolation:
Open
3. Bypass:
Closed
4.
Mirorange Bypass: Closed (microrange active)
Open (microrange inactive)
Figure 3.
molbox RFM
Internal Pneumatic Schematic – MICRORANGE OPTION RPT ACTIVE/INACTIVE
Содержание molbox RFM
Страница 10: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S ...
Страница 142: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 132 N NO OT TE ES S ...
Страница 158: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 148 N NO OT TE ES S ...
Страница 164: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 154 N NO OT TE ES S ...
Страница 170: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 160 N NO OT TE ES S ...