molbox™ RFM™ OPERATION AND MAINTENANCE MANUAL
© 1998-2007 DH Instruments, a Fluke Company
Page 58
If molbox RFM is able to communicate with an RPMx on its COM2 port, the display is:
1.
Real time reading (without head correction) of
upstream (1, Hi) RPT in unit of measure on
line 2.
2.
Real time reading (without head correction) of
downstream (2, Lo) RPT in unit of measure on
line 2.
3.
Real time reading of the RPMx connected to
molbox RFM COM2 to provide the value of
AutoZ
P
offset
.
96.772 kPaa 96.778
Pstd,0:96.7752
Observe the pressure outputs verify that they are stable. A 10 to 15 minute wait,
after venting, is recommended before running AutoZ. When ready, press
[ENT]
to cause AutoZ to run. molbox RFM logs both RPT readings and calculates a
new AutoZ offset value. The display is:
1. Current/previous value of
P
offset
for the
upstream (1, Hi) RPT.
2. Current/previous value of
P
offset
for the
downstream (2, Lo) RPT.
3.
New value of
P
offset
for the upstream (1, Hi) RPT
for the AutoZ that was just run.
4.
New value of
P
offset
for the downstream (2, Lo)
RPT for the AutoZ that was just run.
Old: 0.0 Pa 0.0
New: 3.7 Pa 2.6
Press
[ENT]
to activate the new values of
P
offset
or
[ESC]
to start over.
The value of P
offset
is always displayed and entered in Pascal (Pa).
3.4.4.5
<5BPR> (molbloc-S OPERATION ONLY)
PURPOSE
To quickly measure the molbloc-S upstream and downstream pressure and
calculate and display the BPR (back pressure ratio) when molbox RFM is in a
molbloc-S BPR mode which would not otherwise measure the BPR.
The <5BPR> menu selection is only present during molbloc-S operation.
PRINCIPAL
molbox RFM uses the back pressure ratio, or BPR (the ratio of the molbloc-S
downstream absolute pressure to the upstream absolute pressure) to determine
whether the flow through the throat of the molbloc-S venturi nozzle is critical and
flow measurements within predictable uncertainty limits can be made with
molbloc-S (see Section 3.1.2). When operating molbox RFM with molbloc-S, the
user may select different BPR monitoring modes (see Section 3.6.9). The BPR
function allows the user to measure and display BPR directly at any time
regardless of the current BPR mode.
Содержание molbox RFM
Страница 10: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S ...
Страница 142: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 132 N NO OT TE ES S ...
Страница 158: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 148 N NO OT TE ES S ...
Страница 164: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 154 N NO OT TE ES S ...
Страница 170: ...molbox RFM OPERATION AND MAINTENANCE MANUAL 1998 2007 DH Instruments a Fluke Company Page 160 N NO OT TE ES S ...