Table 8. Leveled Sine Wave Function Verification: Amplitude (cont.)
Nominal Value
(p-p)
Frequency
Measured Value
(p-p)
Deviation (mV)
1-Year Spec.
(mV)
40.0 mV
50 kHz
1.000
50.0 mV
50 kHz
1.200
100.0 mV
50 kHz
2.200
200.0 mV
50 kHz
4.200
400.0 mV
50 kHz
8.200
500.0 mV
50 kHz
10.200
1.3 V
50 kHz
26.200
2.0 V
50 kHz
40.200
5.5 V
50 kHz
110.200
Table 9. Leveled Sine Wave Function Verification: Flatness
Nominal Value
(p-p)
Frequency
Measured Value
(p-p)
Deviation (mV)
1-Year Spec.
(mV)
5.0 mV
500 kHZ
0.18
5.0 mV
1 MHz
0.18
5.0 mV
1 MHz
0.18
5.0 mV
2 MHz
0.18
5.0 mV
5 MHz
0.18
5.0 mV
10 MHz
0.18
5.0 mV
20 MHz
0.18
5.0 mV
50 MHz
0.18
5.0 mV
100 MHz
0.18
5.0 mV
125 MHz
0.20
5.0 mV
160 MHz
0.20
5.0 mV
200 MHz
0.20
10.0 mV
500 kHZ
0.25
10.0 mV
1 MHz
0.25
10.0 mV
1 MHz
0.25
10.0 mV
2 MHz
0.25
10.0 mV
5 MHz
0.25
10.0 mV
10 MHz
0.25
10.0 mV
20 MHz
0.25
10.0 mV
50 MHz
0.25