Scope Calibration Option
Detailed Specifications
3
Edge Function
Edge Characteristics into 50
Ω
Specification, 1-Year, tcal
±
5
°
C
Amplitude
Range
4.5 mV to 2.75 V
±
(2 % of 300
μ
V)
Resolution 4
digits
Adjustment Range
±
10 % around each sequence value
(indicated below)
Sequence
5 mV, 10 mV, 25 mV, 50 mV, 100 mV,
250 mV, 500 mV, 1 V, 2.5 V
Other Edge Characteristics
Frequency Range
900 Hz to 1.1 MHz
±
(5 ppm of s15 mHz)
Rise Time
<1 ns
Within 10 ns
<(3 % of 3 mV)
10 to 30 ns
<(1 % of 3 mV)
Leading Edge Aberrations
After 30 ns
<(0.5 % of 3 mV)
Typical Duty Cycle
45 % to 55 %
Leveled Sine Wave Function
Leveled Sine Wave
Frequency Range
Characteristics into 50
Ω
50 kHz Reference
50 kHz to 100 MHz
100 to 200 MHz
[1]
Amplitude Characteristics
Range (p-p)
5 mV to 5.5 V
Resolution
<100 mV: 3 digits
≥
100 mV: 4 digits
Adjustment Range
Continuously adjustable
Specification, 1-Year,
tcal
±
5
°
C
±
(2 % of 300
μ
V)
±
(3.5 % of 400
μ
V)
±
(4 % of 400
μ
V)
Flatness (relative to 50 kHz)
Not applicable
±
(1.5 % of 200
μ
V)
±
(2.0 % of 200
μ
V)
Short-term Stability
≤
1 %
[1]
Frequency Characteristics
Resolution 10
Hz
10 kHz
[2]
10 kHz
Specification, 1-Year,
tcal
±
5
°
C
±
5 ppm
±
5 ppm
±
5 ppm
Distortion Characteristics
2
nd
Harmonic
≤
-33 dBc
3
rd
and higher Harmonics
≤
-38 dBc
[1]
Within one hour after reference amplitude setting, provided temperature varies no more than
±
5
°
C.
[2]
At frequencies below 120 kHz, the resolution in 10 Hz. For frequencies between 120 kHz and 999.9 kHz, the resolution is 100 Hz.
Time Marker Function
Time Marker into 50
Ω
5 s to 50 ms
20 ms to 100 ns
50 ns to 20 ns
10 ns
5 ns to 2 ns
Specification at cardinal
points, 1-Year,
tcal
±
5
°
C
[2]
±
(50 + t*1500) ppm
[1]
±
5 ppm
±
5 ppm
±
5 ppm
±
5 ppm
Wave Shape
spike or square
spike, square, or
sq20%
spike or square
square or sine
sine
Typical Output Level
>1 V p-p
>1 V p-p
>1 V p-p
>1 V p-p
>1 V p-p
Sequence
5-2-1 from 5 s to 2 ns (e.g., 500 ms, 200 ms, 100 ms)
Adjustment Range
At least
±
10 % around each sequence value indicated above.
Amplitude Resolution
4 digits
[1]
t is the time in seconds.
[2]
Away from the cardinal points, add
±
50 ppm.