5080A/SC
Users Manual
4
Trigger Signal for the Time Marker Function
Time Marker Period
Division Ratio
Amplitude into 50
Ω
(p-p)
Typical Rise Time
5 to 35 ms
off, /1
≥
1 V p-p
≤
2 ns
34.99 ms to 750 ns
off, /1, /10, /100
≥
1 V p-p
≤
2 ns
749.9 to 10 ns
off, /10, /100
≥
1 V p-p
≤
2 ns
9.99 to 2 ns
off, /100
≥
1 V p-p
≤
2 ns
Trigger Signal for the Edge Function
Edge Signal Frequency
Division Ratio
Amplitude into 50
Ω
(p-p)
Typical Rise Time
900 Hz to 1.1 MHz
off, /1
≥
1 V p-p
≤
2 ns
Oscilloscope Connections
Using the cable supplied with the Oscilloscope Calibration Option, attach the SCOPE
connector on the 5080A to one of the channel connectors on your oscilloscope (see
Figure 1).
To use the external trigger, attach the TRIG OUT connector on the 5080A to the external
trigger connection on your oscilloscope. To use the external trigger and view its signal
with the calibration signal, attach the TRIG OUT connector to another channel. See your
oscilloscope manual for details on connecting and viewing an external trigger.
gjm001.eps
Figure 1. Oscilloscope Connection: Channel and External Trigger
Starting the Oscilloscope Calibration Option
Press
to start the Oscilloscope Calibration Option. The Control Display opens the
Volt menu, shown below, which contains options for calibrating the vertical gain on your
oscilloscope. This is the first of four calibration menus, which you can scroll through by
pressing the softkey under MODE. Each menu is described in detail in this chapter.
gjm014.eps