background image

5080A/SC 

Users Manual 

 

gjm004.eps 

Each item in the V/DIV menu is described below. 

 

V/div

 Changes the scale of the output display by changing the number of volts that 

are represented by each division. The available settings, shown in the figure above, 

are provided in 1-2-5 step increments. Press the softkey under UP to increase the 

volts per division. Press the softkey under DOWN to decrease the volts per division. 

 

# DIV

 Specifies the number of divisions that establish the p-p value of the 

waveform. The value can be adjusted from one to eight divisions. The amount 

denoted by each division is displayed in the V/div field.  Press the softkey under UP 

to increase the signal’s height, and press the softkey under DOWN to decrease it. 

Shortcuts for Setting the Voltage Amplitude 

The 

 and 

 keys step the voltages through cardinal point values of an oscilloscope 

in a 1-2-5 step sequence. For example, if the voltage is 40 mV, then pressing 

 

increases the voltage to the nearest cardinal point, which is 50 mV. Pressing 

  

decreases the voltage to the nearest cardinal point, which is 20 mV. 

Amplitude Calibration Procedure for an Oscilloscope 

This example procedure describes how to use the Volt menu to calibrate the 

oscilloscope’s amplitude gain. During calibration, you will need to set different voltages 

and verify that the gain matches the graticule lines on the oscilloscope according to the 

specifications for your particular oscilloscope. See your oscilloscope manual for the 

recommended calibration settings and appropriate gain values. 
Before you start this procedure, verify that you are running the oscilloscope option in 

riešenia na presné meranie

http://www.elso.sk

Содержание 5080A/SC

Страница 1: ...rights reserved Printed in USA Specifications are subject to change without notice All product names are trademarks of their respective companies 5080A SC Oscilloscope Calibration Option Users Manual...

Страница 2: ...zed service center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB Destination Fluke assu...

Страница 3: ...ltage Amplitude on an Oscilloscope 7 The Volt Function 7 The V DIV Menu 7 Shortcuts for Setting the Voltage Amplitude 8 Amplitude Calibration Procedure for an Oscilloscope 8 Calibrating the Pulse and...

Страница 4: ...5080A SC Users Manual ii rie enia na presn meranie http www elso sk...

Страница 5: ...Load 21 5 Voltage Function Verification DC Voltage into a 50 Load 22 6 Voltage Function Verification DC Voltage into a 1 M Load 22 7 Edge Function Verification 23 8 Leveled Sine Wave Function Verific...

Страница 6: ...5080A SC Users Manual iv rie enia na presn meranie http www elso sk...

Страница 7: ...v List of Figures Figure Title Page 1 Oscilloscope Connection Channel and External Trigger 4 rie enia na presn meranie http www elso sk...

Страница 8: ...5080A SC Users Manual vi rie enia na presn meranie http www elso sk...

Страница 9: ...nction This calibration procedure is similar to the one for verifying the vertical deflection characteristics except that it checks the horizontal axis The menus that implement these functions also in...

Страница 10: ...200 m 40 000 ft maximum Detailed Specifications Voltage Function DC Signal Square Wave Signal Voltage Function Into 50 Into 1 M Into 50 Into 1 M Amplitude Characteristics Range 0 V to 2 2 V 0 V to 33...

Страница 11: ...ess relative to 50 kHz Not applicable 1 5 of output 200 V 2 0 of output 200 V Short term Stability 1 1 Frequency Characteristics Resolution 10 Hz 10 kHz 2 10 kHz Specification 1 Year tcal 5 C 5 ppm 5...

Страница 12: ...e the external trigger attach the TRIG OUT connector on the 5080A to the external trigger connection on your oscilloscope To use the external trigger and view its signal with the calibration signal at...

Страница 13: ...Since oscilloscope calibration requires many adjustments of the output signal the three available methods for changing these settings for oscilloscope calibration are summarized below These methods p...

Страница 14: ...ps 5 Press to remove the cursor from the Output Display and save the new value as the reference value Note If you attempt to use the rotary knob to adjust a value to an amount that is invalid for the...

Страница 15: ...tion menus gjm003 eps Each menu item is described below OUTPUT SCOPE Indicates the location of the signal output If the signal does not appear on the oscilloscope press To disconnect the signal press...

Страница 16: ...Shortcuts for Setting the Voltage Amplitude The and keys step the voltages through cardinal point values of an oscilloscope in a 1 2 5 step sequence For example if the voltage is 40 mV then pressing i...

Страница 17: ...mount specified for the calibration settings for your oscilloscope This example shows the gain at 30 mV to be 6 divisions at 5 mV per division gl006i bmp 4 Change the voltage to the next value recomme...

Страница 18: ...Indicates you are in Edge mode Use the softkey to change modes and open the corresponding menus for the other three oscilloscope calibration modes Pulse Response Calibration Procedure for an Oscillos...

Страница 19: ...oscilloscope exhibits the proper rise time and pulse aberration characteristics 7 Remove the input signal by pressing The Leveled Sine Wave Function The Leveled Sine Wave Levsine function uses a level...

Страница 20: ...ortcuts for Setting the Frequency and Voltage Three options are available for controlling the sine wave settings SET TO LAST F toggles between the last frequency used and the reference frequency of 50...

Страница 21: ...sweep rate lets you watch the frequency change very slowly After a fast sweep you may want to pinpoint a certain frequency with a slow sweep over a subset of your previous frequency range RANGE The so...

Страница 22: ...the Sweep menu appears on the Control Display as shown below gl031i eps 4 You can let the signal sweep through the entire range or you can halt the sweep if you need to record the frequency at a certa...

Страница 23: ...y six divisions To fine tune the voltage press to bring a cursor into the Output Display move the cursor with the key and turn the rotary knob to adjust the value See Fine Tuning Values earlier in thi...

Страница 24: ...is available through the Marker menu lets you calibrate the timing response of your oscilloscope To access the Marker menu press the softkey under MODE until marker appears gjm008 eps Each option in t...

Страница 25: ...point which is 500 s Time Base Marker Calibration Procedure for an Oscilloscope This sample procedure uses the Time Marker function to check the horizontal deflection time base of your oscilloscope S...

Страница 26: ...the 5080A Operators Manual Overlapped Commands Commands that require additional time to execute are called overlapped commands because they can overlap the next command before completing execution To...

Страница 27: ...scope Marker mode Programs the period to 1 ms output at the SCOPE BNC standby if from OFF or previously in standby Example SCOPE VOLT OUT 2 V 0 Hz dc voltage 2 V SCOPE VOLT OUT 4 V 1 kHz ac voltage 4...

Страница 28: ...points The verification test points are provided here as a guide when re verification is desired Table 3 Voltage Function Verification AC Voltage into a 1 M ohm Load Nominal Value p p Frequency Measur...

Страница 29: ...cation AC Voltage into a 50 Load Nominal Value p p Frequency Measured Value p p Deviation mV 1 Year Spec mV 5 0 mV 45 Hz 0 2175 5 0 mV 100 Hz 0 2175 5 0 mV 1 kHz 0 2175 10 0 mV 100 Hz 0 235 10 0 mV 1...

Страница 30: ...0 2875 55 0 mV 0 3925 55 0 mV 0 3925 100 0 mV 0 55 100 0 mV 0 55 220 0 mV 0 97 220 0 mV 0 97 250 0 mV 1 075 250 0 mV 1 075 550 0 mV 2 125 550 0 mV 2 125 700 0 mV 2 65 700 0 mV 2 65 2 2 V 7 9 2 2 V 7...

Страница 31: ...1 95 500 0 mV 1 95 3 3 V 11 75 3 3 V 11 75 4 0 V 14 2 4 0 V 14 2 33 0 V 115 7 33 0 V 115 7 Table 7 Edge Function Verification Nominal Value p p Frequency Pulse Response Time nS 1 Year Spec ns 250 0 mV...

Страница 32: ...le 9 Leveled Sine Wave Function Verification Flatness Nominal Value p p Frequency Measured Value p p Deviation mV 1 Year Spec mV 5 0 mV 500 kHZ 0 18 5 0 mV 1 MHz 0 18 5 0 mV 1 MHz 0 18 5 0 mV 2 MHz 0...

Страница 33: ...0 40 0 mV 5 MHz 0 70 40 0 mV 10 MHz 0 70 40 0 mV 20 MHz 0 70 40 0 mV 50 MHz 0 70 40 0 mV 100 MHz 0 70 40 0 mV 125 MHz 0 90 40 0 mV 160 MHz 0 90 40 0 mV 200 MHz 0 90 40 0 mV 220 MHz 0 90 40 0 mV 235 MH...

Страница 34: ...10 400 0 mV 125 MHz 8 10 400 0 mV 160 MHz 8 10 400 0 mV 200 MHz 8 10 1 3 V 500 kHZ 19 60 1 3 V 1 MHz 19 60 1 3 V 1 MHz 19 60 1 3 V 2 MHz 19 60 1 3 V 5 MHz 19 60 1 3 V 10 MHz 19 60 1 3 V 20 MHz 19 60...

Страница 35: ...V 10 MHz 0 0003 MHz 1 3 V 200 MHz 0 0050 MHz Table 11 Marker Generator Function Verification Nominal Interval Measured Interval Deviation 1 Year Spec 5 s 25 13 ms 2 00 s 4 05 ms 1 s 1 03 ms 500 00 ms...

Страница 36: ...enerator Function Verification cont Nominal Interval Measured Interval Deviation 1 Year Spec 100 00 ns 2 500 ps 50 00 ns 1 250 ps 20 00 ns 0 500 ps 10 00 ns 0 250 ps 5 00 ns 0 125 ps 2 00 ns 0 050 ps...

Отзывы: