5080A/SC
Users Manual
22
Table 5. Voltage Function Verification: DC Voltage into a 50
Ω
Load
Nominal Value (dc)
Measured Value (dc)
Deviation (mV)
1-Year Spec. (mV)
0.0 mV
0.2
5.0 mV
0.2175
-5.0 mV
0.2175
10.0 mV
0.235
-10.0 mV
0.235
22.0 mV
0.277
-22.0 mV
0.277
25.0 mV
0.2875
-25.0 mV
0.2875
55.0 mV
0.3925
-55.0 mV
0.3925
100.0 mV
0.55
-100.0 mV
0.55
220.0 mV
0.97
-220.0 mV
0.97
250.0 mV
1.075
-250.0 mV
1.075
550.0 mV
2.125
-550.0 mV
2.125
700.0 mV
2.65
-700.0 mV
2.65
2.2 V
7.9
-2.2 V
7.9
Table 6. Voltage Function Verification: DC Voltage into a 1 M
Ω
Load
Nominal Value (dc)
Measured Value (dc)
Deviation (mV)
1-Year Spec. (mV)
0.0 mV
0.2
5.0 mV
0.2175
-5.0 mV
0.2175
22.0 mV
0.277
-22.0 mV
0.277
25.0 mV
0.2875
-25.0 mV
0.2875