5080A/SC 

Users Manual 

Adjusting Values with the Rotary Knob 

To adjust values in the Output Display using the rotary knob: 
1.

 

Turn the rotary knob. A cursor appears in the output display under the lowest digit 

and begins changing that digit. If you wish to place the cursor in the field without 

changing the digit, press 

 

gjm011.eps 

2.

 

To move the cursor between the voltage and frequency fields, press 

 

gl004i.eps 

3.

 

Use the 

 and 

 keys to move the cursor to the digit you want to change. 

4.

 

Turn the rotary knob to change the value. 

When you use the rotary knob in either Volt mode or Marker mode, the Control 

Display shows the new value’s percentage change from the reference value. This is 

useful for determining the percentage of error on the oscilloscope. You can set the 

reference value to the new value by pressing 

 

gjm013.eps 

5.

 

Press 

 to remove the cursor from the Output Display and save the new value 

as the reference value. 

Note 

If you attempt to use the rotary knob to adjust a value to an amount that is 
invalid for the function you are using, or is outside the value’s range limit, 
the value will not change and the 5080A will beep. If you need to reach a 
different range of values, turn the knob quickly to jump to the new range. 

Using the 

 and 

 

The 

 and 

 keys cause the current value of the signal to jump to a pre-determined 

cardinal value, whose amount is determined by the current function. These keys are 

described in more detail under the descriptions for each function. 

Resetting the Oscilloscope Option 

You can reset all parameters in the 5080A to their default settings at any time during 

front panel operations by pressing the 

 key on the front panel. 

After resetting the 5080A, press 

 to return to the Oscilloscope Calibration Option (the 

Volt menu appears). Press 

 to reconnect the signal output. 

riešenia na presné meranie

http://www.elso.sk

Содержание 5080A/SC

Страница 1: ...rights reserved Printed in USA Specifications are subject to change without notice All product names are trademarks of their respective companies 5080A SC Oscilloscope Calibration Option Users Manual...

Страница 2: ...zed service center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB Destination Fluke assu...

Страница 3: ...ltage Amplitude on an Oscilloscope 7 The Volt Function 7 The V DIV Menu 7 Shortcuts for Setting the Voltage Amplitude 8 Amplitude Calibration Procedure for an Oscilloscope 8 Calibrating the Pulse and...

Страница 4: ...5080A SC Users Manual ii rie enia na presn meranie http www elso sk...

Страница 5: ...Load 21 5 Voltage Function Verification DC Voltage into a 50 Load 22 6 Voltage Function Verification DC Voltage into a 1 M Load 22 7 Edge Function Verification 23 8 Leveled Sine Wave Function Verific...

Страница 6: ...5080A SC Users Manual iv rie enia na presn meranie http www elso sk...

Страница 7: ...v List of Figures Figure Title Page 1 Oscilloscope Connection Channel and External Trigger 4 rie enia na presn meranie http www elso sk...

Страница 8: ...5080A SC Users Manual vi rie enia na presn meranie http www elso sk...

Страница 9: ...nction This calibration procedure is similar to the one for verifying the vertical deflection characteristics except that it checks the horizontal axis The menus that implement these functions also in...

Страница 10: ...200 m 40 000 ft maximum Detailed Specifications Voltage Function DC Signal Square Wave Signal Voltage Function Into 50 Into 1 M Into 50 Into 1 M Amplitude Characteristics Range 0 V to 2 2 V 0 V to 33...

Страница 11: ...ess relative to 50 kHz Not applicable 1 5 of output 200 V 2 0 of output 200 V Short term Stability 1 1 Frequency Characteristics Resolution 10 Hz 10 kHz 2 10 kHz Specification 1 Year tcal 5 C 5 ppm 5...

Страница 12: ...e the external trigger attach the TRIG OUT connector on the 5080A to the external trigger connection on your oscilloscope To use the external trigger and view its signal with the calibration signal at...

Страница 13: ...Since oscilloscope calibration requires many adjustments of the output signal the three available methods for changing these settings for oscilloscope calibration are summarized below These methods p...

Страница 14: ...ps 5 Press to remove the cursor from the Output Display and save the new value as the reference value Note If you attempt to use the rotary knob to adjust a value to an amount that is invalid for the...

Страница 15: ...tion menus gjm003 eps Each menu item is described below OUTPUT SCOPE Indicates the location of the signal output If the signal does not appear on the oscilloscope press To disconnect the signal press...

Страница 16: ...Shortcuts for Setting the Voltage Amplitude The and keys step the voltages through cardinal point values of an oscilloscope in a 1 2 5 step sequence For example if the voltage is 40 mV then pressing i...

Страница 17: ...mount specified for the calibration settings for your oscilloscope This example shows the gain at 30 mV to be 6 divisions at 5 mV per division gl006i bmp 4 Change the voltage to the next value recomme...

Страница 18: ...Indicates you are in Edge mode Use the softkey to change modes and open the corresponding menus for the other three oscilloscope calibration modes Pulse Response Calibration Procedure for an Oscillos...

Страница 19: ...oscilloscope exhibits the proper rise time and pulse aberration characteristics 7 Remove the input signal by pressing The Leveled Sine Wave Function The Leveled Sine Wave Levsine function uses a level...

Страница 20: ...ortcuts for Setting the Frequency and Voltage Three options are available for controlling the sine wave settings SET TO LAST F toggles between the last frequency used and the reference frequency of 50...

Страница 21: ...sweep rate lets you watch the frequency change very slowly After a fast sweep you may want to pinpoint a certain frequency with a slow sweep over a subset of your previous frequency range RANGE The so...

Страница 22: ...the Sweep menu appears on the Control Display as shown below gl031i eps 4 You can let the signal sweep through the entire range or you can halt the sweep if you need to record the frequency at a certa...

Страница 23: ...y six divisions To fine tune the voltage press to bring a cursor into the Output Display move the cursor with the key and turn the rotary knob to adjust the value See Fine Tuning Values earlier in thi...

Страница 24: ...is available through the Marker menu lets you calibrate the timing response of your oscilloscope To access the Marker menu press the softkey under MODE until marker appears gjm008 eps Each option in t...

Страница 25: ...point which is 500 s Time Base Marker Calibration Procedure for an Oscilloscope This sample procedure uses the Time Marker function to check the horizontal deflection time base of your oscilloscope S...

Страница 26: ...the 5080A Operators Manual Overlapped Commands Commands that require additional time to execute are called overlapped commands because they can overlap the next command before completing execution To...

Страница 27: ...scope Marker mode Programs the period to 1 ms output at the SCOPE BNC standby if from OFF or previously in standby Example SCOPE VOLT OUT 2 V 0 Hz dc voltage 2 V SCOPE VOLT OUT 4 V 1 kHz ac voltage 4...

Страница 28: ...points The verification test points are provided here as a guide when re verification is desired Table 3 Voltage Function Verification AC Voltage into a 1 M ohm Load Nominal Value p p Frequency Measur...

Страница 29: ...cation AC Voltage into a 50 Load Nominal Value p p Frequency Measured Value p p Deviation mV 1 Year Spec mV 5 0 mV 45 Hz 0 2175 5 0 mV 100 Hz 0 2175 5 0 mV 1 kHz 0 2175 10 0 mV 100 Hz 0 235 10 0 mV 1...

Страница 30: ...0 2875 55 0 mV 0 3925 55 0 mV 0 3925 100 0 mV 0 55 100 0 mV 0 55 220 0 mV 0 97 220 0 mV 0 97 250 0 mV 1 075 250 0 mV 1 075 550 0 mV 2 125 550 0 mV 2 125 700 0 mV 2 65 700 0 mV 2 65 2 2 V 7 9 2 2 V 7...

Страница 31: ...1 95 500 0 mV 1 95 3 3 V 11 75 3 3 V 11 75 4 0 V 14 2 4 0 V 14 2 33 0 V 115 7 33 0 V 115 7 Table 7 Edge Function Verification Nominal Value p p Frequency Pulse Response Time nS 1 Year Spec ns 250 0 mV...

Страница 32: ...le 9 Leveled Sine Wave Function Verification Flatness Nominal Value p p Frequency Measured Value p p Deviation mV 1 Year Spec mV 5 0 mV 500 kHZ 0 18 5 0 mV 1 MHz 0 18 5 0 mV 1 MHz 0 18 5 0 mV 2 MHz 0...

Страница 33: ...0 40 0 mV 5 MHz 0 70 40 0 mV 10 MHz 0 70 40 0 mV 20 MHz 0 70 40 0 mV 50 MHz 0 70 40 0 mV 100 MHz 0 70 40 0 mV 125 MHz 0 90 40 0 mV 160 MHz 0 90 40 0 mV 200 MHz 0 90 40 0 mV 220 MHz 0 90 40 0 mV 235 MH...

Страница 34: ...10 400 0 mV 125 MHz 8 10 400 0 mV 160 MHz 8 10 400 0 mV 200 MHz 8 10 1 3 V 500 kHZ 19 60 1 3 V 1 MHz 19 60 1 3 V 1 MHz 19 60 1 3 V 2 MHz 19 60 1 3 V 5 MHz 19 60 1 3 V 10 MHz 19 60 1 3 V 20 MHz 19 60...

Страница 35: ...V 10 MHz 0 0003 MHz 1 3 V 200 MHz 0 0050 MHz Table 11 Marker Generator Function Verification Nominal Interval Measured Interval Deviation 1 Year Spec 5 s 25 13 ms 2 00 s 4 05 ms 1 s 1 03 ms 500 00 ms...

Страница 36: ...enerator Function Verification cont Nominal Interval Measured Interval Deviation 1 Year Spec 100 00 ns 2 500 ps 50 00 ns 1 250 ps 20 00 ns 0 500 ps 10 00 ns 0 250 ps 5 00 ns 0 125 ps 2 00 ns 0 050 ps...

Отзывы: