General Specifications
All specifications are valid after a warm-up period of 30 minutes, or twice the time since last warmed up, to a maximum of
30 minutes. For example, if the 5080A has been turned off for 5 minutes, the warm-up period is 10 minutes.
Specifications include stability, temperature coefficient, linearity, line and local regulation, and the traceability of the
external standards used for calibration. It is not necessary to add anything to determine the total specification for the
temperature range indicated.
Specification Confidence Level
........................... 99 %
Warmup Time
........................................................ Twice the time since last warmed up, to a maximum of 30 minutes.
Temperature
Operating ............................................................ 0
°
C to 50
°
C
Calibration (tcal).................................................. 15
°
C to 35
°
C
Storage ............................................................... -20
°
C to +70
°
C
Temperature Coefficient
....................................... Temperature coefficient for temperatures outside tcal
±
5
°
C is 10 % of
the stated specification per
°
C for temperatures in the range of 0
°
C to
35
°
C. Above 35
°
C, the temperature coefficient is 20 % of the stated
specification per
°
C.
Relative Humidity
Operating ............................................................ <80 % to 30
°
C, <70 % to 40
°
C, <40 % to 50
°
C.
Storage ............................................................... <95 %, non-condensing
Altitude
Operating ............................................................ 2,000 m (6,500 ft) maximum
Non-operating ..................................................... 12,200 m (40,000 ft) maximum
Detailed Specifications
Voltage Function
DC Signal
Square Wave Signal
Voltage Function
Into 50
Ω
Into
1
M
Ω
Into
50
Ω
Into
1
M
Ω
Amplitude Characteristics
Range
0 V to
±
2.2 V
0 V to
±
33 V
1.8 mV to 2.2 V p-p
1.8 mV to
105 V p-p
[1]
Resolution
<100 V: 4 digits or 10
μ
V, whichever is greater
≥
100 V: 5 digits
Adjustment Range
Continuous
[1]
Specification, 1-Year, tcal
±
5
°
C
±
(0.35 % of 200
μ
V)
[2] [3]
Sequence
1-2-5 (e.g., 10 mV, 20 mV, 50 mV)
Square Wave Frequency Characteristics
Range
45 Hz to 1 kHz
Specification, 1-Year, tcal
±
5
°
C
±
(50 ppm of s25 mHz)
Typical Aberration within 30
μ
s
from leading edge
<(3 % of 200
μ
V)
[1]
The square wave into 1 M
Ω
is a positive square wave from 1.8 mV to 55 V p-p. From 95 V to 105 V. Its output is a square wave-
like signal that alternates between the negative peak and the positive peak, with the centerline at -10 V. Signals between 55 V and
95 V p-p are not available.
[2]
The uncertainty of 50
Ω
loads does not include the input impedance error of the oscilloscope. Square wave signals below 4.5 mV
p-p have a specification of
±
(0.35 % of 300
μ
V).
[3]
Signals from 95 to 105 V p-p have a specification of 1 % of output in the frequency range 100 Hz to 1 kHz. Typical specification is
3 % of output for 95 V to 105 V p-p signals in the frequency range 45 Hz to 100 Hz.